Felt Pad Assembly - Konica Minolta DRYPRO SIGMA Service Manual

Medical printer
Hide thumbs Also See for DRYPRO SIGMA:
Table of Contents

Advertisement

SERVICE MANUAL

Felt Pad Assembly

Prerequisites:
Removal
Caution
Hot Surface: When the laser imager is initially de-energized, the processor drum rollers are hot. Take care when
working in the area of the processor.
[1] De-energize the laser imager.
Felt Pad Assembly
Screw (1 of 2)
Heat Shield
[2] Loosen the 2 captive thumb screws on the heat shield, and then lift the heat shield open.
[3] Remove the:
• 2 screws that hold the felt pad assembly to the cooling plate
• Felt pad assembly
2-20
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
• Remove:
Top cover
Left cover
Felt Pad Assembly
Captive Thumb Screws
Cooling Plate
9G9002 – 2011-08-15