Power Supply Board

Prerequisites:
T10 Torx driver
Tools:
Removal
ESD
Possible damage from electrostatic discharge.
When installing circuit boards, always use a wrist strap to prevent damage from electrostatic
discharge.
[1] Remove the:
• 2 power supply shield screws
• Power supply shield
[2] Disconnect the input and output connectors, 1 on each side of the power supply board.
H221_6021BC
[3] Remove the:
• 4 screws that secure the power supply board to the standoffs
• Power supply board
Installation
Reverse the steps in the removal procedure.
9G9002 – 2011-08-15
• Close the film cartridge. (Touch the Pause button on the display screen.)
• De-energize the laser imager.
• Disconnect the power cord.
Remove the back
cover.
Power Supply
Board
Connectors
Power Supply
Screw (1 of 4)
Adjustments and Replacements
Power Supply Shield
Shield Screw
(1 of 2)