Scp Setup - Konica Minolta DRYPRO793 Installation Manual

Laser imager
Table of Contents

Advertisement

Ch.3 Setup

3.5 SCP Setup

The procedure for making SCP settings (DICOM information for DRYPRO 793 itself) is described below.
Item
IP ADDRESS
SUBNET MASK
GATEWAY
HOST NAME
PORT NUMBER
AE TITLE
EXPANSION PORT
CHANNEL
DRYPRO Vstage MODEL 793 Installation Manual Ver.1.00 2004.11
Default
192.168.20.161
Sets the DRYPRO 793 IP address.
255.255.255.0
Sets the network sub-net mask.
0.0.0.0
Sets the network gateway.
(Blank)
Sets the DRYPRO 793 host name.
A maximum of 16 alphanumerical characters may be specified.
6000
Sets the DRYPRO 793 DICOM receipt port number.
104
It is not necessary to duplicate settings for the same port number.
0
Do not set all values to "0."
0
Settings may be made within a range of 100 – 65000.
KC_DPRO3_P001
Sets the DRYPRO 793 AE tile.
Alphabetical settings are case sensitive.
NO SELECT
Selects the channel to be used for port expansion in cases where
external devices are connected in the expanded port specification.
<NO SELECT>
<CH01>
~
<CH16>
< 95 >
Descriptions
: Connection according to the basic specification
(Expanded port not used)
: CH01 is set as the expanded port.
: CH16 is set as the expanded port.
p.172 SCP Setup
(Web Mainte. Mode)
p.11 Expanded Port
Specification

Hide quick links: