Measurement In "Auscultation Mode - Omron HBP-1100 Instruction Manual

Professional blood pressure monitor
Hide thumbs Also See for HBP-1100:
Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 1
Measurement in "Auscultation Mode"
In "Auscultation Mode", this device does not measure blood
pressure.Measurement should be performed by a health care
professional using a stethoscope.
The healthcare professional uses a stethoscope to determine SYS
and DIA by means of the auscultation method.
1. Make sure the power is on.
"0" is displayed.
2. Press the [Auscultation] button.
The auscultation icon appears and the
device enters "Auscultation Mode".
3. Press the [START/STOP] button.
Inflation starts. When the cuff
is sufficiently inflated, deflation
automatically starts.
4. At the SYS point that you determine by auscultation, press the
[Auscultation] button.
The first time you press the [Auscultation] button, the SYS value
appears.
5. At the DIA point that you determine by auscultation, press the
[Auscultation] button.
The second time you press the [Auscultation] button, the DIA value
appears and the cuff rapidly deflates.
Manual inflation in "Auscultation Mode"
If the cuff is not inflated sufficiently or you want to re-inflate, you can
inflate the cuff manually.
Hold down the [START/STOP] button during inflation or deflation to
inflate continuously.
"-" appears below the value to indicate that manual inflation is in
progress.
20