ADJUSTMENT FOR MR DUTY

● In case of replacing the main PCB, SWM unit or MR encoder unit, be sure to make this adjustment.
 How to adjust
 ① Make sure that the current and voltage of the connected rated voltage power supply are the set values.
   Then, turn on the rated voltage power supply for the contacts A and F.
 ② Select "1. MR DUTY ADJUSTMENT" in the menu items of the AF-S zoom lens (New) inspection program.
 
 ③ The display to check whether the fixed values are written in EEPROM or not appears. Select a proper item.
 
 ④ According to the instruction on the screen, rotate the MF ring from the infinity direction to the near distance
   direction slowly by hand. Make sure that the waveform on the oscilloscope has duty 50% and then stop the MF
   ring at the near distance end.
C H 1 = 5 V
C H 2 = 5 V
D C 1 0 : 1
D C 1 0 : 1
 ⑤ According to the instruction on the screen, rotate the MF ring from the near distance direction to the infinity
   direction slowly by hand. Make sure that the waveform on the oscilloscope has duty 50% and then stop the
   MF ring at the infinity end.
 
 Note:If the waveform from infinity to near distance and vice versa does not have duty 50%, perform again
 "INSPECTION AND ADJUSTMENT FOR THE WAVEFORM OUTPUT FROM MR ENCODER" in P.L13.
Standard  H:L= 100:206 ~ 206:100(50% ±17.3%)
5 m s / d i v
NORM 200KS/s
- L26 ・ AF-S DX12-24/4G -
JAA78451-R.3596.A
●Setting of oscilloscope
 V/Div(CH1)
 V/Div(CH2)
 Coupling
:DC
:5m Sec
 Time/Div
 Trigger Mode :NORMAL
 Trigger Coupling
:DC
 Trigger Source
:CH1
 Trigger Position
:+4div
 Trigger Type :EDGE
 Trigger Level :2. 5V
INC
:5V
:5V