Use

3
Use
3.1

Prerequisites

The ambient air must be free from alcohol and solvent vapours
(disinfectants).
Keep sufficient distance to antennas of mobile phones and
transmitter sites.
3.1.1

Requirements of the test subject

Observe a waiting period of at least 15 minutes after the last oral
alcohol intake!
Residual alcohol in the mouth can distort the measurement. Even
aromatic beverages (e. g. fruit juice), alcoholic mouth sprays,
medical juices and drops and, belching and vomiting may result in
false data. Also observe a waiting period of at least 15 minutes in
these cases. Rinsing the mouth with water or non-alcoholic drinks
does not replace the waiting time!
The test person should breathe regularly and calmly before the
sample is taken. Avoid breathing in and out quickly through the
mouth.
The test person must be capable of exhaling the minimum breath
volume of 1.2 l (works setting). For the test, the breath flow must
fall within a certain minimum exhalation time (works setting
2 seconds). The volume and the time can be changed via the
menu, if required.
24
3.2

Preparations for use

3.2.1

Inserting or replacing the batteries

CAUTION
!
Explosion hazard!
Do not throw used batteries/rechargeable batteries into a fire
or open by force. Do not recharge alkaline batteries.
Batteries/rechargeable batteries must be disposed of in
accordance with local and national regulations.
Only use the charger indicated in the order list.
Only charge rechargeable NiMH batteries.
NOTICE
i
i
To ensure that the device can be used again immediately
after the alkaline batteries or the rechargeable NiMH batteries
are discharged, it is recommended to have replacement
batteries available.
Replace the batteries within 5 minutes, otherwise the save
time will be reset and must be set again.
1. Release the screw plug on the battery lid (e. g. using a coin) and
remove the battery lid.
2. Insert the batteries and make sure that the polarity is correct.
3. Close the battery lid and tighten the screw plug.
Required batteries:
2 x 1.5 V alkaline type (AAA, LR6, AA) or,
2 rechargeable batteries - 1.2 V NiMH (AAA, LR6, AA)