Error Messages - KERN ALJ 160-4A Operating Manual

Analytical and precision balances
Hide thumbs Also See for ALJ 160-4A:
Table of Contents

Advertisement

12 Error messages

ERR01
ERR02
ERR03
ERR04
ERR05
ERR06
ERR07
ERR08
"UNLOAD":
"CAL But"
72
Weight value instable or zeroing not possible. Check the
environmental conditions.
Adjustment error, e.g. environmental conditions instable
Adjustment error e.g. incorrect adjustment weight}
Piece weight too small/instable
Data transfer not possible, as weighing value is instable. Check
the environmental conditions.
Weighing value in density determining mode instable.
Check the environmental conditions.
Incorrect data reading
(piece counting, density determination,...)
Error at internal adjustment
Weighing range not achieved.
Check position of the weighing plate
Adjust balance
Weighing range exceeded, placed load exceeds the capacity of
the balance. Unload the balance.
Weighing range not reached, e.g. weighing plate not in place
ALJ/ALS/PLJ/PLS-BA-e-1135

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents