5.0 CALIBRATION

Calibration may be required when the balance is initially installed or moved to a new
location. Over time and use, mechanical deviations may occur so it is recommended
that you calibrate you balance regularly for peak performance.
Before calibrating your should make sure you have the correct weights. The weights
should be known to an accuracy that is appropriate for the balance being calibrated,
for example, OIML Class M1 type or ASTM E617 Class 4. If you do not have the
correct weights do not attempt calibration.
Model
CQT 202
Calibration
Weight
100g
Or
200g
1) To calibrate the balance apply power and switch the balance on, whilst the
display is cycling through it's test routine press the [Tare] and [Unit] keys at
the same time, then release after 1 second .
2) The display will show UnLOAD remove any objects from the pan.
3) Press the [Tare] key to enter a zero value.
4) The value last used for calibration will be displayed (see above table). To
select a different calibration weight, press the [Unit] key.
5) Once you have selected the calibration weight to be used press the [Tare]
key.
6) LOAD will be displayed, place the calibration weight on the pan.
7) Once the stable sign is shown press the [Tare] key.
8) If the weight is within 5% of the last calibration value the display will show
PASS and exit the calibration mode. If the calibration is not successful the
display will show FAIL and exit the calibration mode, if the calibration fails try
it again.
CQT 251
CQT 601
200g
100g
Or
Or
400g
200g
Or
600g
CQT 1501
CQT 2601
500g
Or
1000g
1000g
Or
2000g
1500g
13
Adam Equipment 2012
CQT 2000
1000g
Or
Or
2000g
CQT 5000
1000g
Or
3000g
or
5000g