Iso-Tech IDM 98-III Instruction Manual page 22

Digital multimeters
Hide thumbs Also See for IDM 98-III:
Table of Contents

Advertisement

IDM 98-III / 99-III English
(10) Temperature (For IDM99-III only)
Range
-40.0°C ~ 400°C
-40.0°F ~ 752°F
Input Protection : 1000V
* Does not include accuracy of the thermocouple probe.
* Accuracy specification assumes surrounding temperature stable to
± 1°C. For surrounding temperature changes of ± 5°C, rated accuracy
applies after 2 hours.
(11) AC+DC Voltage
Note: When measuring AC+DC, the AC signal may contain a DC level, the AC+DC True
RMS value might not be accurate if the peak voltage of AC+DC value is over full scale.
Range
6.000V
60.00V
600.0V
1000V
Additional specification are same as voltage function.
(12) AC+DC mV
Range
60.00mV
600.0mV
Additional specification are same as voltage function.
IDM99-III
Resolution
0.1°C
0.1°F
or 1000V
rms
DC
AC
IDM98-III/IDM99-III
Resolution
IDM98-III Accuracy
0.001V
0.01V
0.1V
1V
IDM98-III/IDM99-III
Resolution
IDM98-III Accuracy
0.01mV
0.1mV
± (1.0% + 10d)
± (1.0% + 18d)
± (2.5% + 5d)
± (2.5% + 5d)
Accuracy
IDM99-III Accuracy
± (2% + 5d)
IDM99-III Accuracy
± (2% + 10d)