System Calibration - AccuMac AM8010 User Manual

Precision thermometer
Hide thumbs Also See for AM8010:
Table of Contents

Advertisement

AM8010 Precision Thermometer User's Guide
should display:
SAVE Rs
100.0038
Menu = No
And the second line should be blinking. Press
save the new Rs. Press the
After the calibration, turn off power of AM8010 and change
back the Mode Switch to Measurement Mode.

7.2 System Calibration

System Calibration is based on the Instrument Calibration. It is
to integrate a particular PRT sensor with AM8010 after
Instrument Calibration to form a temperature measurement
system. This system is to be calibrated at certain temperature
points to calculate sensor's coefficients based on AM8010's
readings. These coefficients are then saved in AM8010 after
calibration to achieve better measurement performance as a
system.
The advantage of System Calibration is that it can achieve
higher accuracy with lower cost because the sensor doesn't
need additional calibrations before System Calibration.
The following sections illustrate System Calibration procedure
with two types of sensors "IPRT" and "PRT".
7.2.1 Calibration with IPRT (high precision 100 ohm or
industrial 100 ohm Platinum Resistance
Thermometer with α=0.00385)