AccuMac AM1950 User Manual

AccuMac AM1950 User Manual

Standard platinum resistance thermometer

Advertisement

Quick Links

AM1960/1950 Standard Platinum
Resistance Thermometer User's Guide
Please download the latest version at www.accumac.com
Rev. 201609V3

Advertisement

Table of Contents
loading

Summary of Contents for AccuMac AM1950

  • Page 1 AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide Please download the latest version at www.accumac.com Rev. 201609V3...
  • Page 2: Table Of Contents

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide Table of Contents Before you start ---- Warnings & Cautions ................ 2 Introduction ........................ 3 Main Application ....................3 Main Features ..................... 3 Calibrations ......................3 Specifications ......................4 Specifications ...................... 4 General Operations ....................5 Connecting to the readout device ................
  • Page 3: Before You Start

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide Before you start ---- Warnings & Cautions  Warnings: Follow these guidelines to avoid personal injury: 1. Only use this instrument in the manufacture specified temperature range. 2. The handle of this instrument can become hot when it is used to measure high temperatures for extended periods of time.
  • Page 4: Introduction

     Quartz sheathed  Great primary or secondary reference standards 1.3 Calibrations As a temperature standard, each AM1960/AM1950 SPRT must be calibrated. It is highly recommended that the calibration is traceable to a recognized national standards laboratory. Manufacturer recommends calibration interval to be one year. In...
  • Page 5: Specifications

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide 2 Specifications 2.1 Specifications 1950: -200 °C to 500 °C Temperature Range 1960: -200 °C to 670 °C Nominal 25 Ω W(Ga) >= 1.11807 Resistance Ratio W(Hg) <= 0.844235 Drift at 0.01 °C* 1950 <0.002 °C after 100 hours at 500 °C, <0.004 °C/year typical...
  • Page 6: General Operations

    White Yellow Black Figure1 3.2 Drive Current AccuMac recommends 1mA as drive current to ensure the best measurement. 3.3 Stability of Readings To achieve the best accuracy, allow sufficient time for SPRT to stabilize before taking the readings. 3.4 Immersion Requirements Stem effect can cause measurement errors due to heat lost or gained by the sensing element through the thermometer stem.
  • Page 7: Thermal Emf

    Do not submerge SPRT handle when taking measurement. 3.5 Thermal EMF Each AccuMac SPRT has gone through an annealing process and stability test to minimize the thermal EMF, which is caused by either impurities of sensing element or temperature differentials at lead wires connection point.
  • Page 8: Care And Handling Guidelines

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide 4 Care and Handling Guidelines 1. DO NOT subject the SPRT to any physical shocks and vibrations. a. When not using the SPRT, keep it in a place that’s not prone to drop, slam, bang, vibration or other strong physical contacts.
  • Page 9: Troubleshooting

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide 5 Troubleshooting 5.1 Troubleshooting If the SPRT functions abnormally, it could be caused by several possible problem conditions that are described in this section. Try the solutions recommended and if the problems are still not solved, contact manufacture for warranty or repair service.
  • Page 10: Limited Warranty & Limitation Of Liability

    AM1960/1950 Standard Platinum Resistance Thermometer User’s Guide 6 Limited Warranty & Limitation of Liability Each product from AccuMac Corporation is warranted to be free from defects in material and workmanship under normal use and service. The warranty period is 1 year for the Platinum Resistance Thermometer.