AM8010 Precision Thermometer User's Guide
* CALIB MENU
CALIBRATION
2. Press
Rs =
99.9999
Rt = 99.9888
3. The second line shows the internal reference Rs value
(value may be different from meter to meter). The third
line shows the current external standard resistor
resistance reading Rt = 99.9888 ohm, which is not equal
to the real external standard resistor value 100.0038 ohm.
The following Instrument Calibration procedures are to
adjust the Rs value to make Rs value equal to 100.0038
ohm and save the new Rs to the instrument:
4. Press
or
position. Press
position during the adjustments.
5. After the adjustment, the second line value should be
100.0038 ohm.
Rs =
100.0038
Rt = 100.0038
6. Wait for Rt and Rs to be stable ( the status sign should
blink around 10times),then press
Enter = Yes
key,the display will show :
OHM
OHM
key to adjust the Rs value at the cursor
key to right shift the cursor for a new
OHM
OHM
MENU
pad,the LCD
25/33