Display Explanations ................. 9 International Warranty Program. For warranty information, you can con- How to Use the POWER Source ............10 tact the manufacturer, Rossmax International Ltd. Applying the Cuff ................14 Measurement Procedures ............... 17 Attention: Consult the accompanying documents. Please read this Setting Deflation Rate ..............
If you have any problems, please contact the store or the doc- dicated. tor from whom you purchased this unit or please contact Rossmax • Do not connect the air tube or the cuff to other equipment International Ltd.
General advice: Name/Function of Each Part • Do not place or put anything on this unit. • Do not drop this unit. • Turn off power to the unit and unplug the AC adaptor from the Charging electric outlet before moving the unit. LCD Display •...
Real Fuzzy Measuring Technology Display Explanations Display: This unit uses the oscillometric method to detect your blood pressure. Before the cuff starts inflating, the device will establish a baseline cuff pressure equivalent to the air pressure. This unit will determine the ap- propriate inflation level based on pressure oscillations, followed by cuff Memory Zones deflation.
Warning: How to Use the POWER Source • Do not use this unit in places where inflammable gas, such as How to Use the AC Adaptor highly inflammable anesthetic, may be generated or in a high 1. Connect the AC adapter with the AC adapter jack in the back of the pressure oxygen room or an oxygen tent.
Page 7
Battery life: • You can use the unit for approximately six hundred measurements with one charge. • Approximate life of battery pack is two years. However the battery pack life from each charging may be shortened depending on the state of using. Charging time: •...
1. Select cuff according to arm size: pointing your palm, and the artery be in OK range mark over your main artery. Rossmax Cuff size Arm circumference Large Adult Cuff (L) 34~46 cm (13.4”~18.1”) 4. The edge of the cuff should be at approximately 1.5 to 2.5 cm...
Measurement Procedures Here are a few helpful tips to help you obtain more accurate readings: • Blood pressure changes with every heartbeat and is in constant fluc- tuation throughout the day. • Blood pressure recording can be affected by the position of the user, his or her physiological condition and other factors.
This monitor will re-inflate automatically to approximately 220 Recalling Values from Memory mmHg if the system detects that your body needs more pressure to 1. Press the Memory key to view the last previously stored measure- measure your blood pressure. ment.
Troubleshooting Specifications Measurement Method : Oscillometric If any abnormality should arise during use, please check the following Measurement Range : Pressure: 30~260mmHg; points. Pulse: 40~199 beats/ minute EE / Measurement Error: Make sure the L-plug is securely connected Pressure Sensor : Semi conductor to the air socket and measure again.
EMC guidance and manufacturer’s declaration EMC guidance and manufacturer’s declaration Guidance and manufacturer’s declaration-electromagnetic emissions Where P is the maximum output power rating of the transmitter in The AC1000f is intended for use in the electromagnetic environment specified below. watts (W) according to the transmitter manufacturer and d is the The customer or the user of the AC1000f should assure that it is used in such an environment.