Rossmax AK150 F Manual

Rossmax AK150 F Manual

Blood pressure monitor
Hide thumbs Also See for AK150 F:

Advertisement

Quick Links

MODEL: AK150 F
• Superior Accuracy
• Real Fuzzy Technology
• 90 Autimatic Memories
• Easy One Touch Operation
• Easy to Read LCD Display

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AK150 F and is the answer not in the manual?

Questions and answers

Summary of Contents for Rossmax AK150 F

  • Page 1 MODEL: AK150 F • Superior Accuracy • Real Fuzzy Technology • 90 Autimatic Memories • Easy One Touch Operation • Easy to Read LCD Display...
  • Page 2: Name/Function Of Each Part

    This unit is protected against manufacturing defects by an established Inter- national Warranty Program. For warranty information, you can contact the manufacturer, Rossmax International Ltd. or your local distributors. Attention: Consult the accompanying documents. Please read this manual carefully before use. For specific information on your own blood pressure, contact your physician.
  • Page 3: Real Fuzzy Measuring Technology

    3. Real Fuzzy Measuring Technology This unit uses the oscillometric method to detect your blood pressure. Be- fore the cuff starts inflating, the device will establish a baseline cuff pressure equivalent to the air pressure. This unit will determine the appropriate inflation level based on pressure oscillations, followed by cuff deflation. During the deflation, the device will detect the amplitude and slope of the pressure oscillations and thereby determine for you the systolic blood pres- sure, diastolic blood pressure, and pulse.
  • Page 4 5. Blood Pressure Standard The National High Blood Pressure Educa- tion Program Coordinating Committee has developed a blood pressure standard, clas- sifying blood pressure ranges into 4 stages. (Ref. The Seventh Report of the Joint Na- tional Committee on Prevention, Detection, Evaluation, and Treatment of High Blood Pressure-Complete Report JNC-7, 2004).
  • Page 5: Installing Batteries

    Memory Mark: Shows the number of stored measurements Pulse Mark: Shows the pulse rate per minute Weak Battery Mark: Appears when batteries should be replaced Measurement Error: Make sure the L-plug is securely connected to the air socket and measure again quietly. Wrap the cuff correctly and keep arm steady during measurement.
  • Page 6 You need to replace the batteries when: 1. low battery icon appears on display. 2. the ON/OFF/START key is pressed and nothing appears on display. 1. Batteries are hazardous waste. Do not dispose them together with the household garbage. 2. There are no user serviceable parts inside. Batteries or damage from old batteries are not covered by warranty.
  • Page 7 3. Center the tube over the middle of the arm. Press the hook and loop material together securely. Allow room for 2 fingers to fit between the cuff and your arm. Position the artery mark ( ) over the main artery (on the inside of your arm).
  • Page 8 9. Measurement Procedures Important Notes: Here are a few helpful tips to help you obtain more accurate readings: • Blood pressure changes with every heartbeat and is in constant fluctuation throughout the day. • Blood pressure recording can be affected by the position of the user, his or her physiological condition and other factors.For greatest accuracy, wait one hour after exercising, bathing, eating, drinking beverages with alcohol or caffeine, or smoking to measure blood pressure.
  • Page 9 off the power. If no key is pressed, the unit will shut off automatically in 1 minute. This monitor will re-inflate automatically to approx- imately 220 mmHg if the system detects that your body needs more pressure to measure your blood pressure. Note: 1. This monitor automatically switches off approximately 1 minute after last key operation.
  • Page 10: Troubleshooting

    5. If the unit is stored near freezing, allow it to acclimate at room temperature before use. 6. This unit is not field serviceable. You should not use any tool to open the device nor should you attempt to adjust anything inside the device. If you have any problems, please contact the store or the doctor from whom you purchased this unit or please contact Rossmax International Ltd.
  • Page 11: Specifications

    7. For users diagnosed with common arrhythmia (atrial or ventricular premature beats or atrial fibrillation), diabetes, poor circulation of blood, kidney problems, or for users suffered from stroke, or for unconscious users, the device may have difficulty in determining the proper blood pressure. 8. To stop operation at any time, press the ON/OFF/START key, and the air in the cuff will be rapidly exhausted.
  • Page 12 Storage Environment : -10°C~60°C (14°F~140°F); 10%~90% RH max DC Power Source : DC 6V four R06 (AA) Batteries Dimensions : 105 (L) X 140 (W) X 74 (H) mm Weight : 375g (G.W.) (w/o Batteries) Arm circumference : Adult: 24~36 cm (9.4”~14.2”) Limited Users : Adult users : Type BF Device and cuff are designed to provide special protection against electrical shocks.
  • Page 13 Guidance and manufacturer’s declaration-electromagnetic immunity The AK150f is intended for use in the electromagnetic environment specified below. The customer or the user of the AK150f should assure that is used in such and environment. Immunity test IEC 60601 test level Compliance level Electromagnetic environment-guidance Conducted RF IEC...

Table of Contents