Rossmax AC1000f User Manual
Hide thumbs Also See for AC1000f:

Advertisement

AC1000f
Model:
Blood Pressure
Monitor
www.rossmax.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the AC1000f and is the answer not in the manual?

Questions and answers

Summary of Contents for Rossmax AC1000f

  • Page 1 AC1000f Model: Blood Pressure Monitor www.rossmax.com...
  • Page 2: Table Of Contents

    EMC guidance and manufacturer’s declaration ......30-32 vice. If you have any problems, please contact the store or the doc- tor from whom you purchased this unit or please contact Rossmax International Ltd.
  • Page 3: Notes On Safety

    7. As a common issue for all blood pressure monitors using the oscil- used outside temperature and humidity ranges specified in Specifi- lometric measurement function, the device may have difficulty in de- cations. termining the proper blood pressure for users diagnosed with diabe- 17.
  • Page 4: Name/Function Of Each Part

    • Do not use this unit in the presence of flammable gas or anesthetics Name/Function of Each Part or in a high pressure oxygen room or oxygen tent. • Do not use the battery pack for devices other than for this unit. •...
  • Page 5: Parr (Pulse Arrhythmia) Technology

    Monitor Installation - Trolley (Optional) PARR (Pulse Arrhythmia) Technology Placing Protection Jelly Pulse Arrhythmia (PARR) technology specifically detects the existence of onto the Monitor pulse arrhythmia, including atrial fibrillation (AF, AFib), Atrial and / or Ventricular Premature Contractions (PC), Tachycardia (TACH), and Brad- ycardia (BRAD).
  • Page 6: Premature Contraction Detection (Pc)

    Atrial fibrillation most often initially occurs with temporary periods of Note: It is strongly recommended, that you consult your physician, if arrhythmia and may progress to a permanent state of this disorder in either the PC icon occurs newly for several times, or, if your PC is the course of time.
  • Page 7: Pulse Arrhythmia Detection (Arr)

    Pulse Arrhythmia Detection (ARR) Real Fuzzy Measuring Technology Once the occurrence of pulse arrhythmia has been detected in the This unit uses the oscillometric method to detect your blood pressure. course of your blood pressure measurement, the icon ARR is displayed. Before the cuff starts inflating, the device will establish a baseline cuff In the case, that the found pulse arrhythmia can be specified by the pressure equivalent to the air pressure.
  • Page 8: Display Explanations

    Loose Cuff Detection Display Explanations If the cuff was applied too loosely, it may cause unreliable measurement results or measurements can fail to start. The “Cuff Wrap Detection” can help to de- 1. Arrhythmia Detection termine if the cuff is wrapped snugly enough. The specified icon appears (ARR) once a “loosen cuff”...
  • Page 9: How To Use The Power Source

    Output: 7.5V 1.5A Caution: (Ø3.8 / Ø1.35) Brand: ROSSMAX Manufacurer: HON-KWANG • Do not short-circuit the polarities of the optional battery pack with a metal object such as wire. • When using the battery pack and connect the AC adaptor, the unit •...
  • Page 10 How to Use the POWER Source How to Use the POWER Source Battery life: • You can use the unit for approximately six hundred measurements with one charge. • Approximate life of battery pack is two years. However the battery pack life from each charging may be shortened depending on the state of using.
  • Page 11: Applying The Cuff

    Applying the Cuff Applying the Cuff 3. Place the cuff over the bare up- 1. Select cuff according to arm size: per arm, wrap it with the tube Rossmax pointing your palm, and the ar- Cuff size circumference INDEX must tery mark over your main artery.
  • Page 12: Measurement Procedures

    Applying the Cuff Measurement Procedures Here are a few helpful tips to help you obtain more accurate readings: AUTO Mode: MANUAL • Blood pressure changes with every heartbeat and is in constant fluc- without Mode: tuation throughout the day. Stethoscope with Stethoscope •...
  • Page 13 Measurement Procedures Measurement Procedures Korotkoff sound of Systolic pressure 3. Auto Mode • In Auto mode, mark appears on the display. • Press the START key, the monitor will automatically inflate the cuff slowly to start measurement. Korotkoff sound of •...
  • Page 14: Setting Deflation Rate

    Setting Deflation Rate Recalling Values from Memory In the Manual mode, select the deflation rate of 2.5 mmHg/sec, 4.5 1. Press the Memory key to view the last previously stored measure- mmHg/sec, 6.5 mmHg/sec by pressing the key on the back of ment.
  • Page 15: Troubleshooting

    Troubleshooting Specifications If any abnormality should arise during use, please check the following Measurement Method : Oscillometric points. Measurement Range : Pressure: 30~260mmHg; Pulse: 40~199 beats/ minute Symptoms Check Points Correction Pressure Sensor : Semi conductor Charge the battery Have the batteries run Accuracy : Pressure: ±3mmHg;...
  • Page 16: Emc Guidance And Manufacturer's Declaration

    To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the AC1000f is used exceeds NOTE: UT is the a.c. mains voltage prior to application of the test level.
  • Page 17 Blood Pressure Log Blood Pressure Log...
  • Page 18 Blood Pressure Log Blood Pressure Log...
  • Page 19 Batteries, cuff and accessories are not included. Opening or altering the instrument invalidates the guarantee. The guarantee does not cover damage, accidents or non-compliance with the instruction manual. Please contact your local seller/dealer or www.rossmax.com. Customer Name: ________________________________________________ Address: ________________________________________________________...