D
5302 R
RYSTAR
EFERENCE MANUAL
Press the Confirm key to select 'Installation from CF-card'.
3
Before the procedure starts, the system suggests you make a back-up copy of the
current configuration. It is highly recommended to do this, so you can always
restore it in case something should go wrong.
The following screen is displayed:
•
If you want to back up the current configuration on an external CF-card, select 'Ok,
make backup', followed by the Confirm key. The printer will perform the
save configuration procedure'
saved!).
•
If you do not want to back up the current configuration, press the Up/Down arrow
keys to select 'SKIP backup', followed by the Confirm key.
After the back-up is finished (or if you have chosen to skip the back-up), the
4
system will proceed with the actual upgrade.
The 'Installation from CF-card' screen is displayed:
Insert the CF-card.
5
Refer to
If no data available on the CF-card, the following message is displayed:
If there are multiple dataset files present on the CF-card, a list is shown.
6
The 'Select dataset' screen is displayed:
Dataset xyz v1.00
Dataset abc v1.00
Dataset def v1.00
etc...
138
Advanced operation (Key-operator mode)
(See page 103. Note: the complete configuration is
'Inserting a CF-card'
on page 25.
'General
2830A EN 20050526