Establishing the daily operating reference density levels
This procedure enables you to establish the base line values for:
Low density
Mid density
High density
The densitometer of the Drystar 5302 is calibrated at installation.
Authorized service personnel should recalibrate the densitometer
annually or after major service or repair.
To establish the daily operating levels, proceed as follows:
Press the Key-operator key to enter the Key-operator mode.
1
Press the down key seven times, followed by the confirm key to select 'Quality
2
Control'.
5 Restore config.
6 Calibration
7 Service Actions
8 Quality Control
9 Installation
A confirmation screen appears:
Do one of the following:
3
Press the Confirm key (YES) to start the QC procedure and proceed with
step 4.
Press the Escape key (NO) to quit.
Define the tray for printing.
4
The 'Select tray' screen appears:
1.Upper tray
2.Lower tray
128
Advanced operation (Key-operator mode)