Page 4-1, add new paragraphs 4-6A and 4-6B to read:
4-6A.
PEN SYSTEM
4-6B.
The pen assembly consists of a drum type
reservoir resting in a pivot mount which moves along
the carriage arm. A rigid capillary tube feedline leads
from the reservoir to the pen point. Because of the
capillary process, rack mounted (vertical) models
write equally well as the table models.
should never be allowed to run dry.
Page 5-3, add new paragraphs 5-18A and 5-18B to read:
5-18A. PEN MAINTENANCE
5-18B. Pen writing failures may be caused by dried
ink, sediment, air bubbles in feed line, or general
neglect.
Dirt and ink sediment cause the most
difficulty. Dried ink forms brittle crust particles which,
together with dirt, build up in the supply line until the
17999-09423
9260-0128
9260-0127
9270-1006
9270-1007
9270-1023
9270-1027
07035-80630
07035-81030
5950-2587
Page 6-7, Table 6-3, Spare Parts, change K401 to:
07035-60140
07035-60120
Page 6-7, Table 6-3, Spare Parts, delete Pen Holder, add:
07035-80630
Page 6-10, Figure 6-2 (Illustration) is changed as shown in Figure 8-5.
Page 6-11, Figure 6-2 (Legend) make the following changes:
12
47
48
54
56
58
60
61
The pen
Syringe (Pen filler)
Ink, Red
Ink, Blue
Chart Paper, Standard (heavy)
Chart Paper, Standard (light)
Chart Paper, Metric (heavy)
Chart Paper, Metric (light)
Pen Assy
Pen Accessory Kit
Maintenance brochure
Solenoid, Pen Lift
Plunger Assy, Solenoid
Pen Assembly
14600722
Spring, Pen Lift
07035-41820
Block, Pen Arm
0703560185
Pen Scale Assy - Blank Scale
07035-60186
Pen Scale Assy - English Scale
07035-60250
Pen Scale Assy - Metric Scale
07035-61120
Pen Holder Assy (all Serial Prefixes
07035-60070
Pen Holder Assy (Serial Prefix 803 and
07035-61410
Pen Lift Assy
07035-60140
Solenoid Assy (Below S/P 826 requires
07035-62210
Plunger, Solenoid Assy
07035-80630
Pen Assy
ink stops flowing. Pen assemblies should be cleaned
thoroughly every two to four weeks. Clean by soaking
in alcohol or hot water. Clogging during operation
may be cleared by one or more of the following steps:
a. Using the furnished syringe, apply air pressure to
the reservoir's ink filler hole.
b. Soak pen assembly in alcohol or hot water.
c. Internally clean tip by inserting the stiff wire
supplied in accessory kit.
d. Prior to extended storage, flush and clean pen.
CAUTION
Page 6-6, Table 6-2, Mechanical, Electric, add 17999-15126,
Pen Cleaning Wire, H-P, 17999-15126.
Page 6-6, Table 6-2, Accessory Kit, Paper, Ink, change to read
as follows:
before 803)
later without disposable pen)
index point 07005-00730)
Care must be taken not to touch pen
tip
with
fingers
substance.
H-P
17999-09473
Gubelman
H-10070/A
Gubelman
L-10070/A
Gubelman
Gubelman
H-P
07035-80630
H-P
07035-81030
H-P
5950-2587
H-P
07035-60140
H-P
07035-60120
H-P
07035-80630
Superior Spring
H-P
H-P
H-P
H-P
H-P
H-P
H-P
H-P
H-P
H-P
or
any
greasy
8-5