Sartorius SIWABBP Operating Instructions Manual page 67

Hide thumbs Also See for SIWABBP:
Table of Contents

Advertisement

Example: 1
Checkweighing samples with a target weight of 1250 g and a tolerance range from -10 g to +30 g (using the factory-set application
parameters)
Settings (different from the factory settings):
Setup: Application 2: Checkweighing
Setup: Printout, PRTPROT: 7.6, then select the menu line items of your choice
O
O
1240
O
1.) B egin input of target and
tolerance values
2.) P lace a sample with the target
weight (in this example,
1250 g) on the platform
3.) Save target value
®
4.) S ignum
3 only:
Enter value for lower limit
(in this example, 1240 kg)
5.) S ave value for the lower limit
1280
O
p
Setp
+
1.250 kg
Min
+
1.240 kg
Max
+
1.280 kg
G#
+
1.256 kg
T
+
0.000 kg
N
+
1.256 kg
Lim
+
0.48 %
W.Diff+
0.006 kg
-------------------
Application 2: Checkweighing O
®
6.) S ignum
3 only:
Enter value for the upper limit
(in this example, 1280 kg)
S ave value for upper limit
7.) Weigh samples
8.) P rint the results
Note: If automatic printout
of results is enabled, you do
not need to press the p key.
The results are printed
automatically.
Target
Minimum
Maximum
Gross weight
Tare weight
Net weight
P ercentage of deviation from
target*
A bsolute deviation from target
* W hen displayed in relation to
target value:
If the weight is lighter than
the lower limit, the display
shows: LL
I f the weight is heavier than
the target, the display shows:
HH
Signum 1-2-3