Preparation
§ Switch on the scale: Press the e key
§ While all segments are lit, press the
) key
§ Select the Setup menu: Press the
k key repeatedly until Setup is
displayed
Setup
1
wp-1
1.9.
1.10.
1.11.
1.12.
1.13.
1.15.
1.16.
1.17.
1.18.
§ Save settings with the ) key and exit Setup
by pressing the ( key repeatedly
1
) = Not available on scales verified for use in legal metrology
2
) = Factory setting for use in legal metrology
* Factory settings
§ Open the Setup menu:
Press the ) key
$ Select weighing platform 1 "WP1:"
Press the ) key or
$ Select interface 1 "COM-1" or interface
2 "unicom" (depending on the
interface): Press the k key
Select weighing platform 2 "WP2:"
Press the ) key
Calibration and Adjustment
1.9.1 *
Ext. calibration/adjustment; default weight
1.9.3
External calibration/adjustment; weight can be selected under menu item 1.18.1
1.9.4
Internal calibration/adjustment (models with built-in motorized calibration weight only)
1.9.8
Setting the preload
1.9.9
Deleting the preload
1.9.10
No function when you press the
Calibration/Adjustment sequence
1.10.1
Calibration with automatic adjustment
1.10.2 *
Calibration with manual adjustment
Zeroing range
1.11.1
1 percent/max. cap.
1.11.2 *
2 percent/max. cap.
Initial zero point range
1.12.1
Factory setting (depending on the model type)
1.12.2
2 percent/max. cap.
1.12.3 *
5 percent/max. cap.
Tare/zero at power on:
1.13.1 *
On
1.13.2
Off, load previous tare values
1.13.3
Only zero at power on
Adjustment prompt
1.15.1 *
Off
1.15.2
Calibration prompt
External calibration
1.16.1 *
Activated
2
1.16.2
)
Blocked
Calibration weight unit
1.17.1
Grams
1.17.2 *
Kilogram
1
1.17.3
Pounds
)
Enter calibration weight
1.18.1
External user calibration weight (enter value, e.g.: 10,000g)
)
J
key
W
flashes on the display
1
)
Operation
Signum 1-2-3
45