EN-38
4.3 Troubleshooting (Cont.)
Symptom
Error 9.5
Error 9.8
Error 53
LOW REF WT
REF WT Err
------

4.4 Service Information

If the Troubleshooting section does not resolve or describe your problem, contact your authorized
Ohaus Service Agent. For Service assistance or technical support in the United States, call toll-free
1-800-526-0659 between 8:00 AM and 5:00 PM EST. An Ohaus product service specialist will
be available to provide assistance. Outside the USA, please visit our web site, www.ohaus.com to
locate the office nearest you.

4.5 Parts

Description
AC adapters
Output: 12 VAC 500 mA
• US 120V/60Hz
• Euro 230V/50Hz
• UK
230V/50Hz
• Australia 230V/50Hz
Output: 9 VDC 500 mA
• Japan 100V/50Hz
Possible Cause
• Production calibration
not present.
• user calibration data not
present. (Required for
LFT ON only)
• EEPROM checksum error.
• Average piece weight too
small. (Warning)
• Reference weight too small
The weight on the pan is
too small to define a valid
reference weight.
• Busy (tare, zero, printing)
Adventurer Pro
Remedy
• Return balance for service.
• Calibrate balance.
• Cycle power on, off. I f
balance fails to operate,
return for service.
• See section 3.5.3.
• Increase sample size.
• Wait until completion.
Ohaus Part Number
46001724+46001776
46001724+46001774
46001724+46001777
46001724+46001775
46001724+46001776