3.7 Balance Settings

Refer to Section 3.4.2 to enter and navigate the menus.

3.7.1 Calibration

Adventurer Pro balances offer a choice of five calibration methods: Internal Calibration (InCAL
Span Calibration, Linearity Calibration, Calibration Test
• InCal
InCal
InCal
InCal
InCal
TM
- For models equipped with Internal calibration (InCAL
Span
Span
• Span
Span
Span
- Span calibration ensures that the balance reads correctly within
• Linearity
Linearity
Linearity
Linearity - Linearity calibration minimizes deviation between actual and displayed
Linearity
• Cal Test
Cal Test
Cal Test - Calibration test allows the stored calibration data to be tested against the
Cal Test
Cal Test
Cal Adj
Cal Adj
• Cal Adj
Cal Adj
Cal Adj - For models equipped with InCal
Notes: Calibration may be locked out to prevent unauthorized personnel from changing calibration.
if calibration has been locked out, you can only access Cal Test on InCal
Before beginning calibration, have masses available. Masses required for calibration are
listed in the table in section 3.7. 2. Default Span Calibration Points are shown in Bold
print.
balance is accomplished by an internal mass.
specifications using two weight values: zero and a weight value between
25% and 100% of the balance's capacity.
weights within the balance's weighing range. Three weight values are
used: zero, a weight value at midpoint of the balance's weighing range,
and a weight value at or near the balance's specified capacity.
current mass being used for the test.
calibration.
and Calibration Adjust.
TM
, allows adjustment of the internal
TM
Adventurer Pro
TM
), calibration of the
models.
TM
TM
),