Installation Of Mtr-G85 - Sanyo MDF-C8V Service Manual

Ultra-low temperature freezer
Hide thumbs Also See for MDF-C8V:
Table of Contents

Advertisement

TEMPERATURE RECORDER (OPTION)

Installation of MTR-G85

1. Remove 4 screws (A, B, C, D) on the side of the
front panel to take off the front panel. Then remove 4
screws on the cover for the recorder mounting space
and take off the cover (Fig. 1).
2. Fix the temperature recorder to the front panel with
fixture and screws enclosed with the recorder.
Note
Before fixing the temperature recorder, connect the
recorder connector with the connector on the right
back side of the control panel (Fig. 2)
3. Lead the recorder sensor to the chamber through
the access port at the bottom of the chamber.
Note:
The port is covered with a thermal insulation and
rubber cap.
Remove those covers before passing
the recorder sensor.
4. Fix the recorder sensor to the inside of the sensor
cover (MTR-C8) by using the enclosed clips and
screws and seal the access port on the chamber side
with a silicon. (Fig. 3)
33
Front panel
A
B
Recorder sensor
Sensor cover
66
C
D
Cover
Fig. 1
Connector
Recorder
Fig. 2
Silicon
Fig. 3