F16:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F16" displayed to revert to PV display.
F17:
<Purpose> Non-volatile memory initialization, model code change
<Operation> Service code should be input in F06 prior to use this mode.
F18:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F18" displayed to revert to PV display.
F19:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F19" displayed to revert to PV display.
F20:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F20" displayed to revert to PV display.
F21:
<Purpose> Serial communication ID setting
<Operation> Input F21 and press SET key to display "000" (initial).
F22:
<Purpose> Serial communication mode setting
<Operation> input F22 and press SET key to display "000" (initial value)
F23:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F23" displayed to revert to PV display.
F24:
<Purpose> To control remote alarm output
<Operation> Input F24 and Press SET key to display "000" (initial).
Buzzer tone is emitted if service code is not input in F06.
Input F17 and press SET key to display "001". Change the value
with
key and
key. Press SET key to store and revert to PV
display.
Model code '003': MDF-C8V
Buzzer tone is emitted if service code is not input in F06.
Buzzer tone is emitted if service code is not input in F06.
Buzzer tone is emitted if service code is not input in F06.
Change the value in the range "001"~"255" with
Change the value with
Press SET key to store the value and revert to PV display.
rd
Control mode (the 3
digit)
0: Local (initial)
1: Remote
nd
Baud rate (the 2
digit)
0: 2400bps (initial)
1: 4800bps
2: 9600bps
Note) You cannot be changed SV if control mode is set in "Remote".
Buzzer tone is emitted if service code is not input in F06.
Change to the desired value with
Press SET key to store the value and revert to PV display.
000: Remote alarm is ON even if the buzzer stops beeping with BZ
key pressed.
001: Remote alarm is OFF when the buzzer stops beeping with BZ
key pressed.
19
key and
key.
key and
key and
key.
key.