F25:
<Purpose> Alarm auto recovery time setting
<Operation> Input F25 and press SET key to display "030" (initial).
F28:
<Purpose> To display delay time of permission of measuring running rate
<Operation> Service code should be input in F06 prior to use this mode.
F29:
<Purpose> To display delay time of permission of measuring running rate
<Operation> Service code should be input in F06 prior to use this mode.
F30:
<Purpose> ROM version is displayed
<Operation> Service code should be input in F06 prior to use this mode.
F31:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F31" displayed to revert to PV display.
F32:
<Purpose> To notify accumulation time of condensing fan motor
<Operation> Service code should be input in F06 prior to use this mode.
F33:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F33" displayed to revert to PV display.
F34~F39
<Operation> Input F34~F39 and press SET key.
8.
Differential (The point of compressor ON and OFF )
COMP ON:
COMP OFF:
9.
Offset value
PV+1.0
is the offset value to adjust the difference between temperature of temp. sensor
and the center of chamber.
Change to the desired value with
Press SET key to store the value and revert to PV display.
000: Auto recovery OFF
010: Recovers after 10min. passed
020: Recovers after 20min. passed
030: Recovers after 30min. passed
(initial)
(2hrs timer; 000~120 min)
Input F28 and press SET key to display F28 and 'xxx' (present count
value) alternately. Press SET key to revert to PV.
(8hrs timer; 000~480 min)
Input F29 and press SET key to display F29 and 'xxx' (present count
value) alternately. Press SET key to revert to PV.
Input F30 and press SET key to display alternately F30 with "X.XX"
(present ROM version).
Press SET key to revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
Input F32 and press SET key to display '000' (Initial setting).
000: Not notify
001: Notify
Set your desired setting with
Press SET key to store the value and revert to PV display.
Buzzer tone is emitted if service code is not input in F06.
Unused
SV +0.4
SV -0.4
20
key and
key.
040: Recovers after 40min. passed
050: Recovers after 50min. passed
060: Recovers after 60min. passed
key and
key pressed.