7.
Function mode
F00
Automatically revert to PV display
F01
SV
H
F02
SV
(low temp. alarm SV) setting
L
F03
Battery accumulating time display
F04
Automatically revert to PV display
F05
Compressor delay time setting
F06
Service code input (code: 384)
F07 *
Temperature Zero Adjustment
F08 *
Automatically revert to PV display
F09 *
Factory test mode ...... Unavailable
F10 *
Factory test mode ...... Unavailable
F11 *
Factory test mode ...... Unavailable
F12 *
Display of temperature of temp. sensor
F13 *
Automatically revert to PV display
F14 *
Display of temperature of Comp. sensor
F15 *
Automatically revert to PV display
F16 *
Automatically revert to PV display
F17 *
Model code setting (Initialization of non-volatile ROM and memory)
F18 *
Automatically revert to PV display
F19 *
Automatically revert to PV display
F20 *
Automatically revert to PV display
F21
Communication ID set
F22
Communication mode set
F23 *
Automatically revert to PV display
F24
Remote alarm terminal output
F25
Ring-back time set
F26 *
"000" is displayed...Unavailable
F27 *
"000" is displayed...Unavailable
F28 *
Display of delay time of permission for measuring running rate (2 hrs timer)
F29 *
Display of delay time of permission for measuring running rate (8 hrs timer)
F30 *
ROM version is displayed
F31 *
Automatically revert to PV display
F32 *
Notice for condensing fan motor accumulating time
F33 *
Automatically revert to PV display
F34~F39
Unused
Setting
In PV display, keep pressing
procedure:
Input the desired function code with
Press SET key to be function mode available.
* ... You should input service code (384) in F06 beforehand.
To cancel service code, input "000" in F06 or turn the power off.
F00:
<Purpose> Simply passing through if entered by mistake.
<Operation> Press SET key in "F00" displayed to revert to PV display.
F01:
<Purpose> SV
<Operation> Input F01 and press SET key to display "010" (initial value).
F02:
<Purpose> SV
<Operation> Input F02 and press SET key to display "-10" (initial value).
F03:
<Purpose> Battery accumulating time is displayed
<Operation> Input F03 and press SET key to display alternately F03 with "00.0"
(high temp. alarm SV) setting
(high temp. alarm SV) setting
H
Set the value in the range "005"~"020" with
to store the value and revert to PV display.
(low temp. alarm SV) setting
L
Set the value in the range "-05"~"-20" with
store the value and revert to PV display.
(in case battery used for a month or less). Press SET key to revert to
PV display.
key over 5seconds to display "F00".
key and
17
key.
key. Press SET key
key. Press SET key to