INSTALLATION
6. The sensor cover inside freezer chamber is removed, and the thermostat sensor part of the back-up
system is inserted into the hole under freezer temperature style sensor. The hole of the neighborhood of
the thermostat sensor part which inserted (A hole does not open an access port. Therefore, make a
hole in the plus driver.) into freezer inside and the rear side is stopped with silicon. Then, a thermostat
sensor is fixed on the sensor mounting plate by using the clip prepared in the sensor mounting plate in
advance. After that, gain a sensor cover inside in the former street. (Fig. 7)
7. There is a access port which inserts the CO
CO
solenoid into the lower corner hole of MDF-UBK. The CO
2
outlet pipe is inserted in that, and a LCO
MDF-UBK with one screw A (M4 x 10 C It is tight.). (Fig. 8) The
mounting plate A is fixed with two screws A (M4 x 10 C is tight.). The
surroundings of the CO
are blocked with silicon. (Fig. 9) The valve cover mounting plate is
fixed by using the screw which sticks in the freezer in advance. It gains
a valve cover after a CO
10)
8. The wiring ass'y DS which it goes to out of the door switch ass'y
is passed through the nylon clip 6N, and MDF-UBK is stopped with
one screw A (M4 x 10 C It is tight.). (Fig. 11)
9. When all harness management is finished, a solenoid valve
cover is fixed with four screws B (M4 x 10 stainless steel (coating
head)) four places. (Fig. 12) (see page 12.)
11
It is blocked with
the silicone.
Thermostat sensor
(A pipe is made a U
form.)
valve outlet pipe of the
2
solenoid valve is fixed on
2
valve outlet pipe which inserted into freezer
2
valve outlet pipe is bent in the bottom. (Fig.
2
chamber
side
Valve outlet pipe
Copper pipe
valve
2
Silicon
Fig.10
84
screw
Sensor cover
screw hole is
upside
Fig. 7
Fig. 8
Fig. 9
Fig. 11
Fig. 12