Nikon ECLIPSE 50i Instructions Manual page 71

Hide thumbs Also See for ECLIPSE 50i:
Table of Contents

Advertisement

11
Inserting Filter Cubes and Light Shield (when the Epi-fl
attachment is mounted)
(1) Remove the cover on the front left of the
Epi-fl attachment.
(2) Insert the filter cube.
The filter cubes listed below cannot be
inserted directly into the Epi-fl attachment
filter bay. The internal spacer must be
removed or reversed, as described on P. 57.
• UV-2A
• UV-2B
• DAPI
• FITC
• GFP-L
• GFP-B
• TRITC
• Tx-Red
(3) Insert a nameplate into the position with the
same address as the one indicated on the
filter cube select knob on the right side of the
microscope.
(4) Turn the filter cube select knob and insert a
filter cube into the remaining open bay.
(5) Replace the cover.
Chapter 4 Assembly
For clarity, the eyepiece tube is not shown here.
68
Filter cube bay
Filter cube
Spacer
Nameplate
Address indication
Filter cube select knob