Nikon eclipse E200 Instructions Manual page 8

Hide thumbs Also See for eclipse E200:
Table of Contents

Advertisement

WARNING
1. Intended use of the equipment
This microscope is intended mainly for use in microscopic observation of cells and tissue fixed on the
slide using diascopic (transmitted) and episcopic (reflected) illumination.
It is designed for the main purposes of experimentation and observation, in hospitals or other
laboratories, of such cells and tissue within the fields of pathology, cytology.
2. Do not disassemble.
Disassembly may cause malfunction and/or electrical shock, and will lead to the forfeiture of all claims
against warranty. Do not disassemble any part other than those described in this manual. If you
experience any problem with the microscope, notify your nearest Nikon representative.
3. Check the Input Voltage.
The input voltage is indicated in two places at the rear of the microscope: on the nameplate and above
the AC inlet. Confirm that these input voltage indications correspond to the voltage provided in your
region. If not, follow one of the instructions below. The use of microscopes with the different input
voltage indications will cause overcurrent and overheating, which may result in fire or severe damage
to the microscope.
• If the voltage indication on the nameplate differs:
Do not plug in the microscope. Contact your nearest Nikon representative.
• If the voltage indication above the AC inlet differs:
Refer to p. 68 and change the input voltage
setting before turning on the power switch.
For the microscope with the nameplate showing
[100/110/120 V ~]:
The voltage can be set to: AC 100 V, 110 V or 120 V.
For the microscope with the nameplate showing
[220/230/240 V ~]:
The voltage can be set to: AC 220 V, 230 V or 240 V.
4. Use the Specified Lamp, Fuse, and Power Cord.
Use the specified lamp and fuse. Use the power cord provided. Using an incorrect lamp, fuse, or power
cord may damage the instrument or cause a fire. (Also see p. 85 on power cord.)
If using an extension cord, only use a cord that includes a protective earth (PE) wire.
• Specified Lamp
Halogen lamp 6V-20W (PHILIPS 7388 or OSRAM HLX64250) or
Halogen lamp 6V-30W (PHILIPS 5761)
• Specified Fuse
250 V, 1A, time-lag low-breaking type, 5×20 miniature fuse ×2
English
Safety Precautions
6
Input Voltage
Indications

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents