Respironics REMstar User Manual page 33

M series
Hide thumbs Also See for REMstar:
Table of Contents

Advertisement

p
RoblEM
The device's display is
erratic.
Device Resets/
Reboots
The device shuts down
and restarts automati-
cally during therapy.
(This is unlikely to
occur.)
REM
M s
U
staR
ERiEs
sER
W
i
h
hy
t
appEnEd
The device or power
supply has been
dropped or mishan-
dled, or the device
or power supply is
in an area with high
Electromagnetic
Interference (EMI)
emissions.
The device comes in-
stalled with trouble-
shooting software
that automatically
monitors perfor-
mance.
M
anUal
W
hat to
Unplug the device and the power supply.
Reapply power to the device. If the problem
continues, relocate the device to an area with
lower EMI emissions (e.g., away from electronic
equipment such as cellular phones, cordless
phones, computers, TVs, electronic games, hair
dryers, etc.).
If the problem still occurs, contact your home
care provider for assistance.
Such a reset poses no danger to the patient
and assures that the patient receives prescribed
therapy throughout the night. If there is a pos-
sibility of damage to the device, the device will
shut down permanently. The product will then
display the following system alert symbol
to indicate that the device must be returned to
the home care provider for service.