Maintenance
3.8

Electronic components

3.8.1

Working on the electrical system

Remove both fuses from the battery
compartment before carrying out any work
on the electronic system.
Remove the fuses as follows:
Unscrew both crosshead screws (A).
Remove the front cover from the battery tray (B).
Remove both fuses (C).
After the work has been done, replace the fuses in the
reverse order.
3.8.2

Replacing the 'DX' controller

Remove the fuses, see § 3.8.1.
Replace the 'DX' controller as follows:
Push the lock (B) in and simultaneously pull the plug (A) out
of the controller.
Unscrew both screws (C) and the washers, using a Phillips
screwdriver PH-2.
Remove the 'DX' controller (D).
Install a (new) controller in the reverse order.
Replace both fuses: the wheelchair is ready for use.