Nonin GO2 Achieve User Manual page 13

Table of Contents

Advertisement

Temperature (Operating)
Storage/Transportation
Device temperature will not exceed 41°C as measured
during a controlled environment test.
Humidity (Operating)
Storage/Transportation
Operating Altitude
Battery Life (Continuous)
Battery Life (Storage)
21
7077-000-02.indd 21-22
+41°F to +104°F (5°C to +40°C)
-22°F to +158°F (-30°C to +70°C)
10% to 90% relative humidity,
non-condensing
10% to 95% relative humidity,
non-condensing
Up to 40,000 feet /12,192 meters
Approximately 2400 spot
checks based on ~21 hours
of operation using one AAA-
size alkaline battery, calculated
at 30 seconds per use.
6 months minimum
Classifications per IEC 60601-1 / CAN/CSA-C22.2 No. 601.1 / UL 60601-1
Degree of Protection
Enclosure Degree of Ingress Protection
Mode of Operation
This equipment complies with International Standard IEC 60601-1-2:2004 for electromagnetic
compatibility for medical electrical equipment and/or systems. This standard is designed to provide
reasonable protection against harmful interference in a typical installation. However, because of the
proliferation of radio-frequency transmitting equipment and other sources of electrical noise in
health care, home, and many other environments, it is possible that high levels of such interference
due to close proximity or strength of a source might disrupt the performance of this device.
This product complies with ISO 10993-1, Biological evaluation of medical devices – Part 1: Evaluation
and testing.
Type BF-Applied Part
IP33
Continuous
22
11/18/2009 2:20:20 PM