Mount the sensor on the wall

The sensor must be mounted either above the head end of the bed or above the long side
of the bed, according to the images below.
In both cases the sensor should be placed above the centre of the bed side.
Above the head end of the bed
If it is possible to install the sensor in both ways, the long side mounting should be pre-
ferred if a lifting pole is mounted on the bed or if the detection sensitivity setting should
be set to detect the patient on the bed edge, see chapter 4, Using the sensor.
Mounting position 1. Above the head end of the bed
B
A
Above the long side of the bed
1a. If there is not a lifting pole mounted on the bed,
the sensor should be placed 180 cm above the floor,
distance A in the figure.
b. If there is a lifting pole mounted on the bed, the
sensor should be placed 200 cm above the floor and
the vertical distance between sensor and lifting pole
must be at least 40 cm, distance B in the figure.
Note! When the black mark on the sensor cord
reaches the floor, the mounting height is 200 cm
and when the red mark reaches the floor, the
mounting height is 180 cm.
2 Place the bed close to the wall with the sensor
placed just above the middle of the bed, according
to the figure to the left.