7. Troubleshouting

Self-learning error function

Red LED is lit and yellow LED is flashing
Possible cause:
- The sensor is covered by an object.
- The bed or sensor has been moved.
- The IR LEDs do not work.
If the sensor detects one of the above mentioned situations, an alarm will be sent auto-
matically. This is indicated by a flashing yellow LED and a solid red LED.
Solution:
- Remove any blocking obstacles from the sensor´s field of view.
- Check if the bed or the sensor has been moved and put the sensor or the
bed back into place.
- Reconfigure the bed zone, see chapter 3, Installation.
- If the problems still exist, please contact your installer.
Red LED is lit and green LED is flashing
The sensor has a built-in function for finding software errors when the sensor is activated.
In order to repair these errors automaticly, the sensor has to be shut down for a couple of
seconds. Since there is a possibility that the patient leaves the bed during these seconds,
an alarm will be sent.
This type of alarm is indicated by a fixed red LED and a flashing green LED.
Deactivate and activate the sensor with the ON/OFF button.