Switching Diagrams - Wespot SecNurse User Manual

Bed occupancy detector
Table of Contents

Advertisement

Switching diagrams

Out of bed mode
Relay 1 is activated when the patient leaves the bed (or after the countdown timer is
started, depending on the sensitivity/timer). If the patient returns to the bed before the
timer has elapsed the sensor will return to its normal mode, no alarm is sent. In case of an
alarm, relay 1 is activated for 2 seconds.
Relay 2 is activated when the patient is leaving the bed and turned off when the patient
returns to bed again.
Relay 1
Relay 2
Power off Power on
Status
Armed mode
Relay 1 is activated when the patient leaves the bed (or after the countdown timer is
started, depending on the sensitivity/timer). If the patient returns to the bed before the
timer has elapsed the sensor will return to its normal mode, no alarm is sent. In case of an
alarm, relay 1 is activated for 2 seconds.
Relay 2 is activated when the sensor is turned on by the remote control (armed mode) and
turned off when the sensor is turned off again with the remote control. For instance as a
sabotage contact.
Relay 1
Relay 2
Power off Power on
Status
Armed
Disarmed
Armed
Disarmed
Leaves
Alarm
bed
Alarm
Leaves
bed
- 13 -
3. Installation
Alarm
Returns
Sent
to Bed
Alarm
Returns
Sent
to Bed