Escort Prism 20401 Reference Manual page 166

Patient monitor
Table of Contents

Advertisement

Pulse Tone: Pitch varies with SpO
ECG Sync: NELLCOR C-LOCK
SpO
(Nellcor MP405)
2
Display: % SpO
, Plethysmographic Waveform or Pulse Amplitude Bar
2
Range: 0 to 100%
Accuracy: ADULT: 0 to 69% unspecified; 70 to 100% ±2 digits; NEO: 0 to 69%
unspecified; 70 to 100% ±3 digits
Pulse Rate Range: 20 to 250 BPM
Pulse Rate Accuracy: ±3 BPM
High Limit Range: 70 to 100%
Low Limit Range: 50 to 99%
Sensors: Neonatal through Adult (NELLCOR)
Pulse Tone: Pitch varies with SpO
ECG Sync: NELLCOR C-LOCK
SpO
(Masimo SET)
2
Display: % SpO
, Plethysmographic Waveform or Pulse Amplitude Bar
2
Range: 1 to 100%
Accuracy:
Calibrated to functional saturation
During NO Motion Conditions:
Saturation (% SpO
ADULT, PED:
NEO:
Pulse Rate: 25 to 240 bpm ±3 digits (Adult, Ped, and Neo)
During Motion Conditions:
Saturation (% SpO
ADULT, PED:
NEO:
Pulse Rate: 25 to 240 bpm ±5 digits (Adult, Ped, and Neo)
Low Perfusion Performance:
>0.02% Pulse Amplitude
Pulse Rate (bpm)±3 digits
High Limit Range: 70 to 100%
Low Limit Range: 50 to 99%
Sensors: Neonatal through Adult (Masimo)
Pulse Tone: Pitch varies with SpO
INVASIVE PRESSURE
Channels: 1, 2, or 3
Range: -40 to 300 mmHg
Accuracy: ±1% or ±1 mmHg
Parameter Display: SYS, DIAS, and MEAN selectable
Pressure Labels: ART, PA, CVP, ICP, RA, LAP, FEM, UA, UV, PRES
Alarm Limit Range:
Systolic:
ADULT
PED
NEO
Diastolic: ADULT
value
2
value
2
):
2
70 to 100% ±2 digits
70 to 100% ±3 digits
):
2
70 to 100% ±3 digits
70 to 100% ±3 digits
Saturation (%SpO
value
2
0-240 mmHg
0-210 mmHg
0-210 mmHg
0-180 mmHg
12-4
)±2 digits and %Transmission > 5%
2

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Prism 20403

Table of Contents