Escort Prism 20401 Reference Manual page 165

Patient monitor
Table of Contents

Advertisement

Range: 20 to 50° C (68 to 122° F)
Alarm Limit Range 20-44° C
Accuracy: ±0.1° C
Probe: YSI 700 or 400, autosensing
Connector: 1/4 inch (0.6 cm) stereo phone jack
®
SURETEMP
TEMPERATURE
Patient Temperature Range: 28.9 to 42.2 ° C (84.0 to 108.0 ° F)
Predictive Temperature Range: 34.5 to 42.2 ° C (94.0 to 108.0 ° F)
Accuracy: ±0.1 ° C (±0.2 ° F)
Response Time: Oral - approx. 4 sec., Rectal - approx. 15 sec., Axillary – continuous
Monitor Mode: 3 minutes
Cord Length: 9 ft. or 16 ft.
NIBP
Measurement Method: Oscillometric
Measurement Range: Systolic: 30 to 275 mmHg ADULT, 30 to 170 mmHg PED,
30 to 140 mmHg NEO
Diastolic: 20 to 240 mmHg ADULT, 20 to 125 mmHg PED, 20 to 100 mmHg NEO
Alarm Limit Range:
Systolic:
ADULT
PED
NEO
Mean:
ADULT
PED
NEO
Diastolic: ADULT
PED
NEO
Pulse Rate: 30 to 250 BPM
Measurement Time: 40 seconds typical
Modes: Manual - initiated via START key; Automatic - STAT, 1, 2, 3, 4, 5, 10, 15, 30, 45
minutes, 1, 2, 3, 4 hrs
1st Cuff Inflation: ADULT: 90 to 220 mmHg; PED: 60 to 140 mmHg selectable,
NEO: 60 to 140 mmHg selectable
Subsequent Inflations: Systolic +30 mmHg
Alarms: SYS, DIAS, MEAN selectable in 5 mmHg steps
System Accuracy: ±5 mmHg mean error, ±8 mmHg SD
Cuff Sizes: Thigh, Large Adult, Adult, Child, Infant, Neo sizes 1, 2, 3, 4
SpO
(Nellcor MP204)
2
Display: % SpO
, Plethysmographic Waveform or Pulse Amplitude Bar
2
Range: 1 to 100%
Accuracy: ADULT: 70 to 100% ±2 digits; 50 to 69% ±3 digits; NEO: 70 to 94% ±3 digits
Pulse Rate Range: 20 to 250 BPM
Pulse Rate Accuracy: ±3 BPM
High Limit Range: 70 to 100%
Low Limit Range: 50 to 99%
Sensors: Neonatal through Adult (NELLCOR)
50-240 mmHg
25-165 mmHg
25-135 mmHg
40-200 mmHg
15-135 mmHg
15-105 mmHg
30-180 mmHg
10-120 mmHg
10- 95 mmHg
12-3