Escort Prism 20401 Reference Manual page 140

Patient monitor
Table of Contents

Advertisement

Table 5: Remote Status
Remote Status
Null Status
Low Battery
Alarm Suspend
NOTE: In the event a low battery condition occurs coincident with
an alarm suspend, the monitor will prioritize the event. Priority will
be given to the low battery signal and the CIS module will transmit
10h.
Table 6. Ports 1, 2, & 3 Backport Waveform Bit Values
MSB
7
6
Port 1
B7
B6
Waveform
0
0
Invalid Selection
0
1
ECG Waveform
1
0
Invalid Selection
1
1
Invalid Selection
Ports 2 and 3 Waveform Bit Values
B5/2
B4/1
B3/0
Waveform
0
0
0
Invalid Selection
0
0
1
BP1 Scale Mode
0
1
0
BP1 Pulse Mode
0
1
1
BP2 Scale Mode
1
0
0
BP2 Pulse Mode
1
0
1
Respiration
1
1
0
SpO
2
1
1
1
CO
2
Hex Value
00h
No activity
10h
Battery voltage is low
70h
Alarm Suspend activated
5
4
3
Port 2
Ports 1 Waveform Bit Values
9-22
Description
LSB
2
1
0
Port 3