Additional Waveforms

As previously stated, the CIS module transmits two type of packet information, a parameter
packet and a waveform packet. The ECG waveform is always transmitted through the serial port
by default. The Clinical Information System may request additional waveforms by issuing the
following command:
<ESC> <R> <Waveform No.>
The CIS Module will respond by sending the additional waveform packet in the previously
described format (for example, <ESC> <W> <Waveform No.>).
IMPORTANT – Remember the differences in transmission information for 110V/60Hz and
220V/50Hz!
It is particularly important that special attention is applied when counting the number of bytes
received, and to properly use the waveform length byte to locate the appropriate waveform data.
The following command will stop all additional waveforms except ECG:
NOTE:
The ECG waveform packet is always sent by default and
cannot be turned off.
[1Bh] [52h] <Waveform No.>
<ESC> <R> <O>
9-23