Noninvasive Blood Pressure (NIBP) Messages

NIBP MESSAGE
ET = MM:SS
AET = MM:SS
AUTO
CUFF = XXX
LOW OSCIL
ARTIFACT
MAX 2 LOW
CHECK CUF

ETCO2 Messages

ETCO
MESSAGE
2
NO SENSOR
WARM UP
FILT OCCL
CO2 ERROR
ICO2
Indicates elapsed time (ET) since last NIBP measurement.
Indicates elapsed time since last NIBP measurement while auto
timer is on.
Monitor is in NIBP AUTO mode.
Displays as the cuff inflates and deflates, giving the pressure in
mmHg.
Oscillation amplitudes are too low to obtain measurement.
Excessive artifact precluded measurement within a 145-second
timeout.
Cuff inflation was too low to take a reading. May indicate either
incorrect operating mode (ADULT, PED, NEO) or a need to
increase the initial inflation.
40-second timeout exceeded for inflation to at least 20 mmHg.
The MediCO2 module has detected that NO FilterLine is
currently connected. This message will remain displayed until a
FilterLine is connected or another CO2 status message is
displayed.
This message is displayed for approximately 5-10 seconds after
the installation of the FilterLine, or after power-up with the
FilterLine already installed.
The MediCO2 module has detected an airway blockage. If
blockage remains for more than 30 seconds, the MediCO2
module will go into standby mode and this message will remain
displayed until FilterLine is disconnected.
This message will be displayed if the CO/CO2 firmware has not
received valid data from the MediCO2 module for more than 30
seconds.
This message will be displayed when the INSPIRED CO2 value
has exceeded the alarm limit set in the CO2 Setup page in deep
configuration or in the Alarm Limits page under the CO2
Parameter menu.
7-4
MEANING
MEANING