Scanner calibration and focus adjustment
condition. If it becomes scratched or damaged, the calibration results will
be incorrect. Keep the target in the envelope provided and store it in a safe
place. If necessary, a new target can be purchased from your authorized
ScanView dealer.

Performing geometric calibration

1 • Using the registration pins for the Bacher Control 2000 system, place
the XY Calibration Target (PN 960-13-002-60) on the mounting plate.
2 • In ColorQuartet, choose "Calibrate Geometry" from the "Calibration"
menu.
Geometric calibration will automatically be carried out,
and a message will appear to tell you whether it was
successfully completed.
3 • Click on "OK" to confirm.

When is focus calibration necessary?

Focus calibration is rarely necessary for the ScanMate F10, but it should be
carried out at the time of installation.
A special Adjustment, Calibration & Test Target (PN 960-13-002-50) is
supplied with the scanner. Keep it in a safe place after initial calibration, in
case it is needed later.
The ScanMate F10 has a very high depth of field, but if you are not
satisfied with the focus, you can use ColorQuartet's Autofocus function to
get the best possible focus for each individual image (see below). If you are
using Autofocus and are still not getting the results you expect, you should
do a focus calibration.
4•2
ScanMate F10 • Operator's Manual