Rightest Blood Glucose Monitoring System GM550. Please read this Rightest GM550 BGMS is intended in vitro diagnostic use only (outside the body ) and coding is manual thoroughly before testing. It provides all information you need to use the product.
Table of Contents Table of Contents About Rightest GM550 System Quality Control “Lo” and “Hi” readings Package of Rightest GM550 System Kit Quality Control Test Rightest GM550 Meter Precaution Rightest GS550 Test Strip Review Your Test Memory Performing a Quality Control Test...
System Kit Package of Rightest GM550 System Kit Your Rightest Blood Glucose Monitoring System GM550 consists of several items. Please identify 2* each item of your system and learn what they're called and how they're used. These items are included in your Rightest Blood Glucose Monitoring System GM550:...
Page 5
Rightest GM550 Meter Rightest GM550 Meter Test Strip Port Insert test strip here until you hear a click. GM550 Battery Cover Slide cover to change Display Window batteries. Show your test result, messages Meter Battery and relative information. Product name...
Rightest GM550 Meter Rightest GM550 Meter Indicates a test result stored in memory Tell when to apply the blood sample Indicates a test result, not joining into the Average Test strip function of this meter Warns when using the strip, not suitable for your...
Page 7
G 550 Test Strip. - Prevent water from entering the meter. Never immerse the meter or hold it under running Rightest GM550 Meter must only be used with Rightest GS550 Test Strip and water. Rightest Control Solution. The application of other Test Strips or control solutions can lead to - The minimum blood sample size for testing is 1.0µL :(...
Page 8
- For detailed information, please refer to the Rightest GS550 Test Strips Package Insert. - If the Rightest GM550 meters and strips are exposed to a high temperature difference, up and toward meter. blood sample. It gradually turns red please wait 30 minutes before measurement.
Meter activation and Battery Change Meter activation and Battery change Your meter comes with two CR2032, 3volt, batteries installed. Two new batteries will provide power to perform about 1000 tests under normal use. You can press the main button or insert a strip to activate your meter.
Setting Up Your Meter- Setting the Time, Volume and Date Setting Up Your Meter- Setting the Time, Volume and Date You can enter Setting Mode by below two ways. 1. Year setting 1. Reload battery With the year format blinking, press the Left or Right button to adjust it. Then After removing the battery, please press the main button for several times until no signal on screen, press the Main button to confirm it.
Page 11
- When you do not do any settings of meter for over 2 minutes, the meter will leave setting move to measurement of unit setting. mode and power off automatically. - Your Rightest GM550 Meter default is set according to your local preference.
Turning On/Off the Meter Turning on /off the Backlight 1. How to turn on the Meter 1. How to turn on the Backlight Press the Main button or Insert a test strip. Keep pressing the Main button for 2 seconds. 2.
Performing a Blood Test Performing a Blood Test 6. Choose a depth of penetration by rotating the top portion of the depth Depth adjustable cap Release button adjustable cap until the setting depth matches the window. Settings are based on skin type " "...
Page 14
Performing a Blood Test Performing a Blood Test 11. A blood drop symbol appears and is accompanied by a 15. Pull off the depth adjustable cap. Without touching the used further beep (if volume is turned on). Now you can apply disposable lancet, stick the lancet tip into the protective the blood sample within 2 minutes.
Alternative Site Testing (AST) Alternative Site Testing (AST) Alternative site testing-palm or forearm blood sampling PRECAUTION - Fingertip test only: 1. Choose Clear Cap and follow step 1~5 on page 24. - Two hours or less after eating 2. Massage the puncture area of palm or forearm for a few - After exercising seconds.
- If your blood glucose result is unusually high or low, or if you question your testing results, If the Rightest GM550 meters and strips are exposed to a high temperature difference, please repeat the test with a new test strip. You can also run a Quality Control Test to check your meter and test strip.
Test Strips Example of Control Solution Range 83-113 mg/dL If Control Solution results are out of range, your Bionime BGMS may not be working properly. Repeat printed on your test strip vial label. 4.6-6.3 mmol/L the Quality Control Test. If your Control Solution results outside the range still exist, do not use the BGMS to test your blood glucose and contact your Bionime Customer service.
10. The Control Solution result appears. Compare your Quality Control Test result to the Control Solution Range printed on the test strip vial label. GM550 3. While the blood drop symbol flashing, press and hold the main button for over 3 seconds until the "...
Recalling Test Result and Average Maintenance Rightest GM550 Meter is able to store 500 test results with time and date automatically. If your Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying meter has stored 500 results, which is the maximum memory of the meter, the newest test result will case if you do not use it.
Page 20
Recalling Test Result and Average Recalling Test Result and Average 14-days average for example; if your current inquiry time is 2008/1/30, then that you must check that you have tested blood glucose in the past 14 days before 1/30, including today. If not, the 14-day average will show no figure.
Er4 - Applied blood volume is insufficient, please repeat the test with a new test strip. Above, if error screen still appears contact your local Bionime Customer Service. NOTE - When meter backlight has no function, you cannot see the backilght, please change your backlight battery.
Page 22
Bionime Customer Service. please wait for 30 minutes before you do the test.
Specification Specification Measurement Technology Oxidase Electrochemical Sensor Hematocrit 30-60% Measurement Calibration Plasma 2 CR2032 batteries Power Supply Sample Capillary whole blood Meter Battery Life About 1000 tests Minimum Sample Volume 1.0 microliters Meter Dimension 90.6 mm x 46 mm x 16.5 mm Measuring Range 10-600 mg/dL 0.6-33.3 mmol/L 53.0 ±5g with batteries...
Limitations Accuracy Rightest GM550 BGMS is not intended for serum or plasma test. The accuracy of the test study of the Rightest™ Blood Glucose Meter was demonstrated by - Inaccurate test results may be obtained at high altitude more than about 3048 meters (10,000 feet) comparing whole blood (plasma equivalent) glucose values on the Rightest™...
The percent ( and number ) of samples for which the difference between the of problems with the Rightest GM550 products, please contact your local Bionime customer Rightest™ meter value and the Olympus value were within the difference range between the Olympus value service.
Please complete and return the enclosed warranty card to your local Bionime affiliate. Lancing device Manufacturer: Bionime Corporation If the Rightest GM550 Meter and strip are exposed to a high temperature difference, please wait 30 694, Renhua Road, Dali Dist., Taichung City 412, Taiwan minutes before measurement.
Need help?