Page 3
The Rightest GM700 BGMS, combines simple instruction and If you have any question or concern, please contact your local Bionime Customer Service or call us at design with accurate testing results to provide a more reliable, user-friendly diabetes management 886 0800 371 688 GMT+08:00, 08:30~17:30.
Page 4
Precaution Precaution - Using appliance in tropical and / or temperate climate. - Before using the Rightest GM700 BGMS to test your blood glucose, read all the instructions in - Keep the test strips, control solution or vial cap away from children. They may cause a choking this user's manual and practice the tests including the quality control test.
Rightest GM700 Meter Performing a Quality Control Test Understanding Control Test Results Rightest GS700 Test Strip Recalling Test Results The Rightest GM700 Auto Code Key Preparing to test Recalling Average Test Results Battery Installation Caring for Your Meter Setting Up the Meter...
Package of Rightest GM700 System Kit Rightest Blood Glucose Monitoring System GM700 Rightest GM700 BGMS consists of several items. Please identify each item of the system and learn the name and function of each item. These items are included in the...
Rightest GM700 Meter Rightest GM700 Meter Up Button ( Test Strip Port For operating the meter, Insert test strip here until you hear a please check details in click. following sections. Auto Code Key Down Button ( For operating the meter,...
Page 8
Rightest GM700 Meter Rightest GM700 Meter Indicates a test result stored in memory Tell when to apply the blood sample Indicates a test result, not joining into the Average Appears after you insert test strip into meter function of this meter...
Rightest GS700 Test Strip Rightest GS700 Test Strip The Rightest GM700 Meter must only be used with Rightest GS700 Test Strips and Rightest Control Solution GC700. The application of other Test Strips or control solutions can lead to incorrect results.
Auto Code Key is designed to automatically detect the Code of strip to make each lot of strips work The Rightest GM700 meter kit comes with one CR2032, 3 volt, battery installed in the meter. well on the meter. Therefore, the Auto Code Key is not necessary to replace every time.
Battery Installation Setting Up the Meter 4. The meter will then do an electronic test immediately. The screen will begin flashing as soon as the You can enter the setting mode by below two ways: battery is installed. 1. Reload battery After removing the battery, please press the main button for several times until no signal on screen, 5.
Page 12
Setting Up the Meter Setting Up the Meter 1. Year setting 4. Time format 12/24H selection With the year format blinking, press the Up or Down button until the current With the time format blinking, press the Up or Down year appears.
- When you do not do any settings of meter for over 2 minutes, the meter will leave setting mode and power off automatically. - Your Rightest GM700 Meter is ready to use at the first time. The meter default is set according to your local preference.
Function Screens and Button Operations Function Screens and Button Operations Switching the Function Screens 4. Non-Averaging set and canceled: You may select the unwanted or doubtful test results, to exclude them from the average 1. Press the " " button to switch function screens including Memory, Average and Time in calculation, after your every test.
Handling The Rightest GS700 Test Strip Handling The Rightest GS700 Test Strip Inserting the Test Strip: Removing the test strip: 1. Hold the test strip between thumb and middle finger with the indication 1. Hold the test strip in the same position as inserted. symbol facing upright.
Preparing for Testing Performing a Blood Test Before Performing a Blood Glucose test, 1. Hold the depth adjustable cap in one hand and hold the hub in the other hand. Bend the cap towards the down prepare the items below: side.
Page 17
Performing a Blood Test Performing a Blood Test 7. Hold the hub in one hand and pull on the plunger in the 13. Touch and hold the drop to the edge of sample entry until other hand. The device will be cocked. Release the plunger, you hear a "...
Performing a Blood Test Alternative Site Testing (AST) 17. Discard the used disposable lancet into an appropriate Alternative site testing-palm or forearm blood sampling puncture-proof or biohazard container. 1. Choose Clear Cap and follow step 1~5 on page 29. Palm 18.
Alternative Site Testing (AST) View Window appearance Make sure your blood sample covers the PRECAUTION whole area of the View Window to get an - DO NOT test on the palm or forearm if you are testing for hypoglycemia (Low blood accurate test result.
View Window appearance Comprehending Test Results and Messages Blood glucose test results are shown on the meters as your preset measurement unit. PRECAUTION If obtained blood glucose results are unusually high or low or results are of a questionable nature, - Place the blood sample only to the sample entry of the test strip.
Please use Control Solution tested with Rightest GM700 BGMS under Control Solution Mode. If the screen. Please repeat your test with by a new test strip. test result is within the Control Solution Range printed on the strip vial label, the Rightest GM700 If you still get “...
Page 22
About Quality Control Testing About Quality Control Testing When should a Quality Control Test Be Performed? PRECAUTION - Whenever checking if the system is working properly. - Each time you open a new bottle of Control Solution, write the discard date on the label. - Whenever practicing testing and checking correct procedure.
Performing a Quality Control Test Performing a Quality Control Test 1. Take one test strip from vial and Re-cap the vial cap immediately. PRECAUTION 2. Insert the test strip with view window, facing up, into test strip port. - Your Control Solution results will not be calculated for average reading but still can be 3.
Your control solution test results should fall within the control solution range. If the results are within The Rightest GM700 Meter is able to store 1000 test results with time and date automatically. If the range, the Rightest GM700 BGMS is working correctly.
3. The quality control solution test result can be recalled from the saved data. The Rightest GM700 meter provides several average test results. View the 1-day,7-days, 14-days, 30-days, 60-days and 90-days average of test results for better blood monitoring of blood monitoring When the data with "...
Caring for Your Meter Error Messages and Trouble Shooting Maintenance Temperature Error Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying 1. In order to get accurate testing, please do testing between 6~44 °C (43~111 °F ).
Page 27
Please, re-insert the unused test strip again after reviewing the steps on how to insert a test strip If one of the above error messages still appears, please contact Bionime correctly on p. 26 or call Customer Service for support. This error message Customer Service.
Error Messages and Trouble Shooting Limitations Rightest GM700 BGMS is not intended for serum or plasma test. CAUTION - Severe dehydration may cause inaccurately low results. A blood sample should ONLY be applied to the test strip after the test strip has been inserted correctly and the Meter's screen is showing the image of the test strip and a blood drop - Inaccurate test results may be obtained at high altitude more than about 3048 meters (10,000 feet) flashing symbol.
Specification Specification Measurement Technology Hematocrit Dehydrogenase Electrochemical Sensor 10~70% Capillary, Venous, Arterial and Neonatal Whole One CR2032 Battery Sample Power Supply Blood 0.75 microliters Minimum Sample Volume About 1000 tests Battery Life Measuring Range 10-600 mg/dL ( 0.6-33.3 mmol/L ) 99 mm x 46 mm x 17.5 mm Meter Dimension 5 seconds...
If you have any questions or concerns, please contact your Bionime authorized representative. We will make every effort to This warranty does not apply to the performance of a Rightest GM700 Meter that has been altered, provide assistance and solutions for you.
Product complied with In Vitro Diagnostic Medical Device Directive 98/79/EC (CE0197) From 100 to 125 mg/dL (5.6 to 6.9 mmol/L) Impaired fasting glucose (pre-diabetes) EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck / Switzerland E-mail: info@bionime.ch 126 mg/dL (7.0 mmol/L) and above on more than Diabetes...