Blood Glucose Monitoring System
GM280B
User's Manual

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the RIGHTEST GM280B and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for BIONIME RIGHTEST GM280B

  • Page 1 Blood Glucose Monitoring System BIONIME CORPORATION No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan Tel: +886 4 2369 2388 BIONIME GmbH GM280B Fax: +886 4 2261 7586 Tramstrasse 16 User’s Manual http: //www.bionime.com 9442 Berneck E-mail: info@bionime.com...
  • Page 3 RIGHTEST Blood Glucose Test Strip GS280 to check that the meter and test strips are working If you have any questions or concerns, please contact your local Bionime Customer Service or together properly and that the test is performing correctly.
  • Page 4 Caution Caution - Before using the RIGHTEST Blood Glucose Monitoring System GM280B to test your blood - Avoid contact with dripping or splashing liquids. - The minimum blood sample size to test using the RIGHTEST Blood Glucose Monitoring System glucose, please read all of the instructions and conduct all of the tests including the quality control GM280B is 0.75 μL: ( test.
  • Page 5 Limitations Limitations - RIGHTEST Blood Glucose Monitoring System GM280B is not intended for serum or plasma - Please refer to the (Alternative site testing (AST)) chapter if you want to test on palm or forearm. testing. - DO NOT test on alternative site (palm, forearm) if you are testing for insulin dose calculations or - Do not use at altitudes greater than 10,000 feet (3,048 meters), it may cause inaccurate test for Continuous Glucose Monitoring (CGM) devices calibration, or hypoglycemia (Low blood glucose).
  • Page 6: Table Of Contents

    Table of Contents Table of Contents About the RIGHTEST Blood Glucose Monitoring System GM280B Understanding Test Results and Messages..................The RIGHTEST Blood Glucose Monitoring System GM280B ............The RIGHTEST Blood Glucose Meter GM280B ................Quality Control The RIGHTEST Blood Glucose Test Strip GS280................About Quality Control Testing......................
  • Page 7: The Rightest Blood Glucose Monitoring System Gm280B

    The RIGHTEST Blood Glucose Monitoring System GM280B The RIGHTEST Blood Glucose Monitoring System GM280B 3 * Contents Your RIGHTEST Blood Glucose Monitoring System GM280B consists of several items. Please identify each item, learn its name and how it is used. Below are the items included in your RIGHTEST Blood Glucose Monitoring System GM280B: 1.
  • Page 8: The Rightest Blood Glucose Meter Gm280B

    The RIGHTEST Blood Glucose Meter GM280B The RIGHTEST Blood Glucose Meter GM280B Test Strip Port Display Window Meter Label Main Button ( Meter Battery Battery Cover Left Button ( Right Button ( 2D barcode with serial number...
  • Page 9 The RIGHTEST Blood Glucose Meter GM280B The RIGHTEST Blood Glucose Meter GM280B Indicates a test result stored in memory Indicates when to apply the blood sample Appears after you insert test strip into meter Indicates the average result Indicates a control solution test result Appears when an error occurs Current date under time mode or testing date Unit of test result...
  • Page 10: The Rightest Blood Glucose Test Strip Gs280

    - For detailed information, please refer to the RIGHTEST Test Strip GS280 insert. - If the RIGHTEST GM280B meters and strips are exposed to a high temperature sample. It gradually turns difference, please wait 30 minutes before measurement.
  • Page 11 Meter Activation and Battery Change Meter Activation and Battery Change Your RIGHTEST Blood Glucose Meter GM280B comes with one CR2032, 3 volt, battery installed. one new battery will provide power to perform approximately 1000 tests under normal use. Press the main button or insert a strip to activate your meter. 1.
  • Page 12 Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume You can enter the Setting Mode two ways. 1. Year setting With the year format blinking, press the Left or Right button until you see the 1.
  • Page 13 Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume 4. Time format 12/24H selection 7. Volume Setting With the time format blinking, press the Left or Right With the volume blinking, press the Left or Right button to turn the volume on or off.
  • Page 14 Turning On / Off the Meter Handling the RIGHTEST Blood Glucose Test Strip GS280 1. How to turn on the RIGHTEST Blood Glucose Meter GM280B How to handle the RIGHTEST Blood Glucose Test Strip GS280. Press the Main button or Insert a test strip. Inserting the RIGHTEST Blood Glucose Test Strip GS280: 1.
  • Page 15 Handling the RIGHTEST Blood Glucose Test Strip GS280 Getting Ready for Testing Removing the RIGHTEST Blood Glucose Test Strip GS280 Before performing a blood glucose test, prepare the items below : 1. Hold the RIGHTEST Blood Glucose Test Strip GS280 as shown.
  • Page 16 Performing a Blood Glucose Test Performing a Blood Glucose Test 1. Hold the depth adjustable cap in one hand and hold the 7. Hold the base in one hand and pull on the plunger with hub in the other hand. Bend the cap towards the down the other hand.
  • Page 17 Performing a Blood Glucose Test Performing a Blood Glucose Test 12. After inserting a test strip, the " " icon will flash on 15. Countdown Mode will begin on the display window. After the screen. Pressing the Right or Left button to choose 5 seconds, your test result will appear.
  • Page 18 Performing a Blood Glucose Test Setting Markers According to actual life status, you can set meal markers with test result. CAUTION You have to perform meal marker setting before testing. - Do not apply your blood drop to the sample port on the strip until you see the " "...
  • Page 19 Search " RIGHTEST CARE " with iOS or Android system. You may start with the function from pairing the glucose meter to App. More instruction of App, please click " Learn and Assist " for " Operation Guide " on the App. For more information, please visit the official website http://www.bionime.com Bluetooth pairing. Data uploading. GM280B...
  • Page 20 Alternative Site Testing Alternative Site Testing Alternative site testing: palm or forearm blood sampling Palm CAUTION - Consult your healthcare professional before sampling from your palm or forearm. Forearm - To perform alternative site testing, install the clear cap for - Alternate site testing should only be done during steady-state time (when glucose is not your RIGHTEST Lancing Device GD500.
  • Page 21 - Do not reuse RIGHTEST Blood Glucose Test Strip GS280. immediately. - If the RIGHTEST GM280B meters and strips are exposed to temperature environments out of range for the meter – 10 °C (50 °F) or above 40 °C (104 °F) or a high temperature Follow local regulations or consult your health care provider for appropriate disposal of used test difference, please wait 30 minutes before measurement.
  • Page 22: About Quality Control Testing

    Normal: 107-145 5.9-8.1 High: 315-426 17.5-23.7 If you still get a " " result, contact your healthcare professional. NOTE - If you want purchase control solutions, please contact your authorized Bionime representative or Bionime customer service (printed on manual cover).
  • Page 23 About Quality Control Testing About Quality Control Testing When should a Quality Control Test Be Performed ? CAUTION - To ensure that your RIGHTEST Blood Glucose Meter GM280 B and RIGHTEST Blood Glucose Test - Each time you open a new bottle of control solution, write the 1st opening date on the Strip GS280 are working properly.
  • Page 24: Performing A Quality Control Test

    Performing a Quality Control Test Performing a Quality Control Test 1. Take one RIGHTEST Blood Glucose Test Strip GS280 from the vial 9. Tightly replace the cap on the RIGHTEST Control Solution GC550 bottle. CLOSE and close the vial cap immediately. 10.
  • Page 25: Recalling Test Results

    Understanding Control Test Results Recalling Test Results Your control solution test results should fall within the control solution range. If the results are within The RIGHTEST Blood Glucose Meter GM280 B is able to automatically store a maximum of 500 test the range, the RIGHTEST Blood Glucose Monitoring System GM280B is working correctly.
  • Page 26: Recalling Average Test Results

    Recalling Average Test Results Recalling Average Test Results The RIGHTEST Blood Glucose Meter GM280 B provides several average test results. View the 1- 5. Quick Searching: day, 7-days, 14-days, 30-days, 60-days and 90-days average test results for better blood glucose To view test results in sequential order automatically, enter the Memory Mode.
  • Page 27: Maintain The Products

    Maintain the Products Maintain the Products The cleaning procedure is to remove dust, blood and body fluid from the surface and should be Cleaning and Disinfecting frequency: at least once a week. performed whenever the meter or lancing device is visibly dirty. Performing the cleaning procedure To clean the meter: once per week is recommended.
  • Page 28: Error Messages And Troubleshooting

    Monitoring System GM280B authorized distributor or contact your local Bionime customer service. 3. If the RIGHTEST GM280B meters and strips are exposed to temperature environments out of range for the meter – 10 °C (50 °F) or above 40 °C (104 °F) or...
  • Page 29 Strip GS280 has been inserted incorrectly more than twice. Please, re-insert the unused test strip again or call Customer Service for support. This error If one of the above error messages still appears, please contact Bionime message can also indicate that you may have used the wrong test strips.
  • Page 30: Specifications

    Specification Specification Measurement Technology Oxidase Electrochemical Sensor Meter Dimension 82 mm x 45 mm x 15.5 mm Measuring Range 10 - 600 mg/dL (0.6 - 33.3 mmol/L) Meter Weight 43 ± 5g with battery Test Time 5 seconds Monitor Display Area 34 mm x 27.5 mm Memory Capacity 500 blood glucose test results with date and time...
  • Page 31: Warranty

    If you have any questions or in case of problems with the products, please contact your local Bionime customer service or email to This warranty does not apply to the performance of a RIGHTEST Blood Glucose Meter GM280B rightest@bionime.com .
  • Page 32: Expected Glucose Values

    Product complies with in vitro Diagnostic Medical Device Directive 98/79/EC. (CE0197) Pre-diabetes (Impaired fasting glucose) From 100 to 125 mg/dL (5.6 to 6.9 mmol/L) EC-Rep: Bionime GmbH, Tramstrasse 16, 9442 Berneck, Switzerland 126 mg/dL (7.0 mmol/L) and above on more Diabetes...