Summary of Contents for BIONIME GM720

  • Page 3 It is recommended that the blood glucose should be monitored regularly. The Rightest blood glucose monitoring system GM720 is accurate and easy to use, and is therefore your reliable assistant for diabetes management.
  • Page 4 They must not be used on more than one person due to the risk of cross-infection. The Rightest blood glucose monitoring system GM720 is manufactured and distributed by Bionime Corporation. If you have any questions or concerns, please contact your local Bionime customer service (see page 73).
  • Page 5: Intended Use

    Intended use The Rightest blood glucose monitoring system GM720 is intended for in-vitro (outside the body) diagnostic use for self testing. The coding of the test strips is automatically detected. The test result is plasma equivalent and achieved by using (a) fresh capillary whole blood that may be drawn from the fingertip, palm or forearm, and (b) venous whole blood samples.
  • Page 6 Intended use For a better understanding of this user’s manual, hereinafter the Rightest blood glucose monitoring system GM720 will be referred to as BGMS; the Rightest blood glucose meter as meter GM720; the Rightest blood glucose test strips GS720 as test strips; the Rightest lancing device GD720 as lancing device;...
  • Page 7 Table of contents First steps Rightest GM720 blood glucose meter Rightest GM720 package contents Edit marker: how to add/change markers of your results Rightest blood glucose meter GM720 First start-up Records: recalling test results How to perform a blood glucose...
  • Page 8 GM720 Rightest GS720 blood glucose test strips Maintenance and cleaning the meter Handling Error messages and trouble shooting Technical specifications Disposing of the Rightest blood glucose meter GM720, blood glucose test strips GS720 and lancets Warranty Customer service Manufacturer...
  • Page 10 Rightest blood glucose meter GM720 Test strip port Insert test strips here until you hear a click. Display Shows test results, messages and all other relevant information. 07.10.2014 Menu Main/confirm button Single click: confirm. Left and right Double click: return to scroll buttons home-screen.
  • Page 46 BGMS. for four months. Your control solution is diluted. Contact your local Bionime The cap of the test strip vial or the control solution customer service (see page 73). has been left open for a long time.
  • Page 55: Alternative Site Testing (Ast)

    Select the clear AST cap to perform a measurement on an alternative site testing. AST cap can be ordered at your local customer service. Please contact your local Bionime customer service (see page 73).
  • Page 62 Precautions Before using the BGMS to test your blood glucose, please read the instructions for use carefully. The meter must only be used with Rightest blood glucose test strips GS720 and Rightest control solution GC700. The use of other test strips or control solutions may lead to incorrect results.
  • Page 63 Precautions The minimum blood sample size for testing is 0.7 µL. 0.7 µL 1.5 µL 2.0 µL 3.0 µL 4.0 µL Please take a minimum of 0.7 µL to perform the test on the BGMS. A blood sample size above 3.0 µL might contaminate the test strip port and the meter. Make sure your blood sample covers the whole area of the view window to get an accurate test result.
  • Page 65: Maintenance And Cleaning The Meter

    Maintenance and cleaning the meter Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying case when not in use. If the meter is dropped or damaged, perform a quality control test (see page 39) before doing a blood glucose test to insure that the meter is still functioning correctly.
  • Page 70: Technical Specifications

    Technical specifications Measurement technology Electrochemical sensor (FAD GDH) Reference device Olympus (Hexokinase) Measurement calibration Plasma Sample Capillary and venous whole blood Minimum sample volume 0.7 μL Coding Autocoding Measuring range 10 – 600 mg/dL / 0.6 – 33.3 mmol/L Test time 5 sec.
  • Page 71: Battery

    Technical specifications Operating temperature 6 ˚C – 44 ˚C Operating relative humidity < 90% Hematocrit 10 – 70 % Power supply 2 CR2032 batteries Meter battery life 600 measurements Meter dimension 71.3mm * 39mm * 14mm Meter weight 50 +/- 5 g Monitor LCD display...
  • Page 73: Warranty

    This warranty does not apply to the performance of a Rightest blood glucose meter GM720 that has been altered, misused, tampered with or abused in any way. This warranty applies only to the original purchaser of the Rightest blood glucose meter GM720.
  • Page 74: Customer Service

    We are keen to provide a comprehensive service to our customers. Please review all the instructions to ensure that you are performing the steps correctly. If you have any ques- tions, or in the event of problems with the Rightest GM720 products, please contact your local Bionime customer service.
  • Page 76: Manufacturer

    GS720, Rightest control solution GC700 Manufacturer: Bionime Corporation, No. 100, Sec. 2, Daqing St., South Dist., Taichung City 40242, Taiwan The products comply with In Vitro Diagnostic Directive 98/79/EC (CE0197). EU Representative: Bionime GmbH, Tramstrasse 16, CH-9442 Berneck, Switzerland...
  • Page 77 Manufacturer Meter storage conditions -10~60˚C(14~140˚F) Test strip storage conditions 4~30˚C (39~86˚F), <90% relative humidity...