Page 3
Venous, arterial, and neonatal blood testing is limited to healthcare professional use only. Rightest Blood Glucose Monitoring System GM782 is manufactured and supported by Bionime Corporation. If you have any questions or concerns, please contact your local Bionime Customer Customer Service or call us at +886-0800-371-688 GMT+08:00, 08:30~17:30.
Page 4
Caution Caution - Before using the Rightest Blood Glucose Monitoring System GM782 to test your blood glucose, - Avoid contact with dripping or splashing liquids. - The minimum blood sample size to test using the Rightest Blood Glucose Monitoring System please read all of the instructions and conduct all of the tests including the quality control test GM782 is 0.75µL: (...
Table of Contents Table of Contents About the Rightest Blood Glucose Monitoring System GM782 Quality Control Rightest Blood Glucose Monitoring System GM782 ............About Quality Control Testing......................Rightest Blood Glucose Meter GM782 ................Performing a Quality Control Test....................... Rightest Blood Glucose Test Strip GS700 ................
Page 6
Rightest Blood Glucose Monitoring System GM782 Rightest Blood Glucose Monitoring System GM782 3 * Contents GM782 Your Rightest Blood Glucose Monitoring System GM782 consists of several items. Please identify each item, learn its name and how it is used. 4 * Below are the items included in your Rightest Blood Glucose Monitoring System GM782:...
Page 7
Rightest Blood Glucose Meter GM782 Rightest Blood Glucose Meter GM782 Test Strip Port GM782 Display Window GM782 Main Button ( Meter Battery Battery Cover Left Button ( Right Button ( 2D barcode with serial number...
Page 8
Rightest Blood Glucose Meter GM782 Rightest Blood Glucose Meter GM782 Indicates a test result stored in memory Indicates when to apply the blood sample Indicates a test result not included into the Appears after you insert test strip into meter average results of the meter Appears when an error occurs Indicates the average result...
Page 9
Rightest Blood Glucose Test Strip GS700 Rightest Blood Glucose Test Strip GS700 Rightest Blood Glucose Meter G 782 can only be used with Rightest Blood Glucose Test Strips GS700 and Rightest Control Solution GC700. The use of other test strips or control solutions can lead to incorrect results.
Page 10
Meter Activation and Battery Change Meter Activation and Battery Change Your Rightest Blood Glucose Meter GM782 comes with one CR2032, 3volt, batteries installed. GM782 GM782 GM782 new batteries will provide power to perform approximately 1000 tests under normal use. Press GM782 the main button or insert a strip to activate your meter.
Page 11
Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume You can enter the Setting Mode two ways. 1. Year setting With the year format blinking , press the Left or Right button until you see the 1.
Page 12
Setting Up Your Meter - Setting the Date, Time and Volume Setting Up Your Meter - Setting the Date, Time and Volume 4. Time format 12/24H selection 7. Volume Setting With the time format blinking, press the Left or Right With the volume blinking, press the Left or Right button to turn the volume on or off.
Turning On / Off the Meter Handling the Rightest Blood Glucose Test Strip GS700 1. How to turn on the Rightest Blood Glucose Meter GM782 How to handle the Rightest Bl ood Glucose Test Strip GS700 Press the Main button or Insert a test strip. Inserting the Rightest Blood Glucose Test Strip...
Handling the Rightest Blood Glucose Test Strip GS700 Getting Ready for Testing Removing the Rightest Blood Glucose Test Strip GS700 Before performing a blood glucose test, prepare the items below : 1. Hold the Rightest Blood Glucose Test Strip GS700 as shown.
Performing a Blood Glucose Test Performing a Blood Glucose Test 1. Hold the depth adjustable cap in one hand and hold the 7. Hold the base in one hand and pull on the plunger with base with the other hand. Bend the cap down. When a the other hand.
Page 16
Performing a Blood Glucose Test Performing a Blood Glucose Test 12. A blood drop icon will appear on the display window 16. Pull off the depth adjustable cap. Without touching the and will be accompanied by a beep (if volume is turned used disposable lancet, insert the lancet tip into the on).
Performing a Blood Glucose Test Setting Markers According to actual life status, you can set meal markers with test result. CAUTION You have to perform meal marker setting immediately after the meter displays the reading of your - Do not apply your blood drop to the sample port on the strip until you see the " "...
Page 18
Upload Data Airplane Mode After the marker setting, Bluetooth will turn itself on and start to pair with your mobile device or PC. Turn on Airplane Mode Once the connection successfully pairs, test results will start uploading to the connected mobile Press on the right or left button for 3 seconds to turn it on.
Alternative Site Testing Alternative Site Testing Alternative site testing: palm or forearm blood sampling Palm CAUTION - Test results may vary if blood samples are taken from different sites or under certain Forearm - To perform alternative site testing, install the clear cap for conditions where glucose levels are changing rapidly such as: following a drink, a meal, your Rightest Lancing Device GD500.
View Window Appearance Understanding Test Results and Messages Make sure your blood sample covers the If your blood glucose result is unusually high or low, or if you question your test results, repeat the whole area of the view window to get an test with a new Rightest Blood Glucose Test Strip GS700.
Page 21
Understanding Test Results and Messages About Quality Control Testing Rightest Blood Glucose Meter GM782 displays results between 10 What is a Quality Control Test ? and 600 mg/dL. To ensure proper monitoring function, it is necessary to regularly perform a quality control test. Use one of the Rightest Control Solutions GC700 when testing your...
Page 22
About Quality Control Testing About Quality Control Testing When should a Quality Control Test Be Performed ? CAUTION - To ensure that your Rightest Blood Glucose Meter GM782 Rightest Blood Glucose Test - Each time you open a new bottle of control solution, write the expiration date on the label. Strip GS700 are working properly.
Page 23
Performing a Quality Control Test Performing a Quality Control Test 1. Take one Rightest Blood Glucose Test Strip GS700 from the vial 9. Tightly replace the cap on the Rightest Control Solution GC700 bottle. CLOSE 10. The control solution result will appear. Compare your Quality Control Test result and close the vial cap immediately.
Page 24
Understanding Control Test Results Recalling Test Results Your control solution test results should fall within the control solution range printed on the test strip Rightest Blood Glucose Meter GM782 is able to automatically store a maximum of 500 test vial label. If the results are within the range, the Rightest Blood Glucose Monitoring System results with time and date.
Page 25
Recalling Average Test Results Recalling Average Test Results Rightest Blood Glucose Meter GM782 provides several average test results. View the 1-day, 5. Quick Searching: 7-days, 14-days, 30-days, 60-days and 90-days average test results for better blood glucose To view test results in sequential order automatically, enter the Memory Mode. Press the Right or monitoring.
Page 26
Error Messages and Troubleshooting Caring for Your Meter Maintenance Temperature Error Keep your meter and test strip free of dust, water or any other liquid. Store the meter in the carrying case when not in use. If your meter is dropped or damaged, perform a quality control test with the 1.
Page 27
If one of the above error messages still appears, please contact Bionime strip correctly on p.23 or call Customer Service for support. This error message Customer Service.
Page 28
Error Messages and Troubleshooting Limitations Rightest Blood Glucose Monitoring System GM782 is not intended for serum or plasma testing. CAUTION - Inaccurate test results may be obtained at altitudes greater than 10,000 feet (3048 meters) above A blood sample should ONLY be applied to the test strip after the test strip has been sea level.
Page 29
Specification Specification Measurement Technology Hematocrit 10-70% Dehydrogenase Electrochemical Sensor Sample Capillary, Venous, Arterial and Neonatal whole blood One CR2032 Battery (3 Voltage) Power Supply Approximately 350 standard tests Minimum Sample Volume 0.75 microliters Battery Life 82 mm x 45 mm x 15.5 mm Measuring Range 10 - 600 mg/dL Meter Dimension...
Page 30
Warranty Customer Service Bionime Corporation warrants that this product will be free from defects in materials and We sincerely like to provide complete, considerate services to our customers. Please review all the workmanship for five years from the date of purchase.
Expected Glucose Values Without Diabetes Federal Communications Commission (FCC) Statement Expected glucose values without diabetes ⁽¹⁾ Federal Communications Commission (FCC) Statement 15.21 You are cautioned that changes or modifications not expressly approved by the part responsible for compliance could void the user's Status Range (mg/dL) authority to operate the equipment.