Appendix A — Electromagnetic
Contents

Electromagnetic Emissions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Electromagnetic Immunity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1

Frequency Separation Distances . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Electromagnetic Emissions
Note:
The 90217 ABP monitor has been tested under laboratory conditions and is suitable
for use in all establishments, including domestic establishments and those directly
connected to the public low-voltage power supply network that supplies buildings
used for domestic purposes. The customer, or user, of the module should ensure that
it is used in such an environment.
Emission Test
RF emissions
CISPR 11
Electromagnetic Immunity
Note:
The ABP monitor is intended for use in the electromagnetic environment specified
below. The customer, or user, of the module should ensure that it is used in such an
environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-2
Compliance
Group 1
The ABP monitor uses RF energy only for internal
Class B
function. Therefore, RF emissions are very low and
are not likely to cause any interference in nearby
electronic equipment
IEC 60601
Compliance
Test Level
Level
±6 kV contact
6 kV contact
±8 kV air
8 kV air
Compatibility
Electromagnetic Environment
Electromagnetic Environment
Floors should be wood, concrete, or
ceramic tile. If floors are covered with
synthetic material, the relative
humidity should be at least 30%.
A-1