Operational Verification - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Ambulatory Blood Pressure Monitor 90217 Service Manual
Hardware Safety Timers Check
1 From the Main Menu, select 2. MAKE CONTACT WITH MONITOR.
2 When the Main Menu appears, select 11. MANOMETER MODE.
3 Manually pump up the system pressure to 165 mmHg ±10.
4 Remove the IR cable and start the timer.
5 Exit the X program and return to DOS by pressing Esc two times.
6 When the valve opens (pressure decreases rapidly), verify that the elapsed time is between 178
and 182 seconds (fault holdoff timer).
7 When the 90217 display changes from showing pressure to showing the software version,
verify that the elapsed time falls between 194 and 198 seconds (watch dog timer).
8 Remove the test setup and restore any 90217 screws previously removed.

Operational Verification

These procedures verify that the 90217 ABP monitor's blood pressure readings are consistent with
design standards.
Note:
These procedures use the DynaTech Nevada CuffLink Blood Pressure Simulator. If
you are using a different simulator, refer to its operator's manual and determine
equivalent tests. Refer to the end of this Maintenance chapter for instructions on using
the CuffLink simulator in the automatic or the manual mode.
Equipment Required
DynaTech Nevada CuffLink Noninvasive Blood Pressure Analyzer (and equivalent) and
associated tubing, manuals, fittings (software version 2.0 or higher is required)
1/8 inch I. D. tubing (P/N 162-0019-00 or equivalent)
Adult cuff (P/N 016-0264-00 or equivalent)
Luer connector (P/N 103-0008-00)
Blood Pressure Simulator Preparation
1 Turn ON the CuffLink simulator and allow it to warm up for a minimum of 15 minutes.
2 Verify that the calibration sticker is current. If it is not, the simulator will need to be calibrated by
the manufacturer or its authorized service facility.
3 Check the zero pressure by selecting ADAMS Adult from SelectBp in the Main Menu.
4 Press ENT.
5 Press F5 to zero the pressure.
6 Press Esc to return to the Main Menu.
4-10