Operational Tests - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Ambulatory Blood Pressure Monitor 90217 Service Manual

Operational Tests

Conduct the following procedures to verify proper operation of the 90217 ABP Monitor.
Equipment Required
90219-02 system
90219-03 ABP Base Station software
Type AA alkaline batteries (3)
Cable, 90217 to PC, P/N 012-0097-xx (greater than -02)
Setup for air leaks (refer to Air Leak Test on page 2-8)
Analog manometer (optional)
Stop watch or equivalent
Battery spring insertion tool, P/N 003-0084-00
Visual Inspection
1 Check the display window and ensure that it is clear (free from scratches, contamination, etc.)
with the words SYS DIA appearing on the left side.
2 Verify that the front panel START/STOP button responds with a clear "snap" when pressed.
3 Inspect inside the battery compartment to ensure that the battery springs are clean and provide
a good electrical contact with the batteries.
4 Install the three AA batteries and verify that the battery door closes and latches properly.