Manifold Kit - P/N 050-0110-Xx - Spacelabs 90217 Service Manual

Ambulatory blood pressure monitor
Hide thumbs Also See for 90217:
Table of Contents

Advertisement

Ambulatory Blood Pressure Monitor 90217 Service Manual

Manifold Kit - P/N 050-0110-xx

Note:
Manifold kit (P/N 050-0110-00) is used for equipment that has a luer lock
connector.
Manifold kit (P/N 050-0110-01) is used for equipment with a quick-disconnect
connector.
To replace the manifold in the 90217 monitor:
1 Disassemble the 90217 ABP Monitor and remove the 672-0171-xx assembly. Separate the
manifold from the rest of the assembly by cutting through the 176-0279-00 four-conductor
flexstrip cable. Refer to Disassembly Procedures on page 4-15.
2 Unsolder the remaining section of flexstrip from the 672-0171-xx assembly and clear the holes
of solder.
3 From the back (non-display) side of the display board, insert the leads from the new flexstrip.
They are pre-attached to the manifold and should fit into the cleared holes in the old assembly.
Do not bend or stress the new flexstrip any more than necessary during replacement.
4 Ensure that the flexstrip is squarely aligned with the display board and fully seated. Solder the
four leads in place and trim excess lead length.
5 Install the 672-0171-xx assembly back into the case.
6 Perform a full monitor calibration. Refer to Calibration Procedures on page 4-4.
7 Reassemble the 90217 monitor.
4-14

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

90217q

Table of Contents